Macherey-Nagel NANOCOLOR VIS II photometer, 320–1100 nm, without Wi-Fi, 919650.1
€4,233.00
% Regular price: €4,980.00 (15% saved)
- Spectrophotometer for 320–1100 nm with reference detector technology
- 10.1″ HD touchscreen with icon-based menu navigation
- Over 200 pre-programmed tests, 99 freely programmable
- Nephelometric turbidity measurement in accordance with ISO 7027 (860 nm)
- GLP-compliant data storage for 5,000 measurement values
- Interfaces: LAN, USB, RS232 – LIMS-compatible
- CE-certified, 2-year warranty
Product information "Macherey-Nagel NANOCOLOR VIS II photometer, 320–1100 nm, without Wi-Fi, 919650.1"
Precise water and wastewater analysis requires reliable measurement technology that is easy to use in day-to-day laboratory work. The NANOCOLOR VIS II photometer from Macherey-Nagel covers the wavelength range from 320–1100 nm and combines high-quality optics with an icon-based, touchscreen menu system for a quick start without the need for lengthy training.
Precise measurement technology for everyday laboratory use
At its core, the instrument utilises monochromator optics and reference detector technology (RDT). Photometric linearity is < 0.5 % (≤ 2 A) or ≤ 1 % (> 2 A), wavelength accuracy is ± 1 nm and the spectral bandwidth is < 4 nm. A complete scan is completed in less than a minute – ensuring reproducible results in daily measurement routines. The instrument is also insensitive to extraneous light and measures via an open cuvette chamber.
Operation, Methods & Connectivity
Operation is via barcode technology and a 10.1″ LED-backlit HD display with a projective-capacitive touchscreen (PCAP). Over 200 pre-programmed tests and 99 freely programmable methods cover absorbance, transmittance, factor, kinetic and scan measurements. Connections to laboratory information management systems (LIMS) are available via LAN, 2× USB (host), 1× USB (function) and RS232; the internal data storage (16 GB Micro-SDHC) can hold 5,000 measured values and 100 spectra or colour measurements. Software updates can be installed free of charge via a USB stick.
Quality Assurance & Compliance
Data storage is GLP-compliant, and internal quality control is carried out via NANOCONTROL NANOCHECK 2.0. In addition, the instrument offers nephelometric turbidity measurement at 860 nm in accordance with ISO 7027 (measuring range 0.1–1000 NTU) – turbidity is thus detected and reported during the measurement itself. The photometer is CE-certified and comes with a 2-year warranty.
Versatile
In addition to standard measurements (absorbance, transmittance, factor), the NANOCOLOR VIS II also covers kinetic measurements, colour measurements and turbidity determinations. The cuvette chamber accommodates round cuvettes (16 mm OD) as well as rectangular cuvettes (10 mm, 20 mm, 40 mm, 50 mm) and is designed for use with NANOCOLOR round and rectangular cuvette tests.
Technical details
| Type | Spectrophotometer with reference detector technology (RDT) |
| Wavelengths | 320–1100 nm |
| Wavelength resolution | 0.1 nm |
| Wavelength accuracy | ± 1 nm |
| Wavelength reproducibility | < 0.5 nm |
| Wavelength selection | Automatic, barcode, manual |
| Wavelength calibration | Automatic |
| Spectral bandwidth | < 4 nm |
| Optics | Monochromator |
| Photometric range | ± 3.0 E in the wavelength range 340 nm–900 nm |
| Photometric linearity | < 0.5% at ≤ 2 A; ≤ 1% at > 2 A |
| Photometric accuracy | 0.005 A at 0.0–0.5 A; 1 % at 0.5–2.0 A |
| Stray light | < 0.1 % |
| Scanning speed | 1 complete scan in less than 1 min |
| Display | 10.1″ LED-backlit HD display, anti-reflective cover glass with projective capacitive touchscreen (PCAP) |
| Operation | Barcode technology, icon-based on-screen menu navigation, touchscreen |
| Measurement modes | Over 200 pre-programmed tests and special methods, 99 freely programmable methods, absorbance, transmittance, factor, kinetics, scan, nephelometric turbidity measurement |
| Pre-programmed tests | NANOCOLOR round cuvette tests and rectangular cuvette tests |
| Cuvette well | Round cuvettes 16 mm OD, rectangular cuvettes 10 mm, 20 mm, 40 mm, 50 mm |
| External light | Insensitive to extraneous light, open shaft |
| Turbidity measurement | Nephelometric turbidity measurement at 860 nm (in accordance with ISO 7027), 0.1–1000 NTU |
| Internal quality control | With NANOCONTROL NANOCHECK 2.0 |
| Data storage | 16 GB Micro SDHC card, 5,000 measured values, 100 spectra or colour measurements, GLP-compliant |
| Interfaces | LAN, 2 x USB (host), 1 x USB (function) and RS232 |
| Update | Free via USB stick |
| Language | DE / EN / FR / ES / PT / PL / HU / NL / CZ / RO / IT |
| Light source | Halogen lamp |
| Power supply | Input 110–240 V AC, output 12 V 3 A DC |
| Operating range | 10–40 °C, max. 80% relative humidity (non-condensing) |
| Dimensions (L x W x H) | 360 x 400 x 110 mm |
| Weight | 4 kg |
| Gross weight (including packaging) | 7272 g |
| Packaging dimensions | 542 x 474 x 276 mm |
| CE-certified | Yes |
| Hazardous substances | No |
| Warranty | 2 years |
| Trademark | NANOCOLOR |
| Contents | Photometer, manual (quick start guide), touch pen, dust cover, mains adaptor with country-specific plugs, USB cable, USB stick, calibration cuvette, certificate and cleaning cloth |