Skip to main content Skip to search Skip to main navigation

Brand Transferpette -8 electronic Multichannel Pipette, 8-Channel, 1-20 µl, European Power Supply, DE-M, 705400

ⓘ  This product description has been translated from German using AI. Despite careful review, we do not guarantee that the translation is error-free; in case of doubt, the German version shall prevail.

Product information "Brand Transferpette -8 electronic Multichannel Pipette, 8-Channel, 1-20 µl, European Power Supply, DE-M, 705400"

Dilution series in the lower two-digit microlitre range call for consistent dosing over many repetitions. The Transferpette -8 electronic with 1-20 µl does this column by column across eight channels.

Battery in multichannel operation

Three cylindrical AAA single cells (3.6 V, 700 mAh, NiMH) form the built-in battery, which drives all channels together and, when fully charged, lasts a whole pipetting day. Charging is done via a power supply for Europe with an input voltage of 100 to 240 V at 50/60 Hz and an output voltage of 6.5 V DC, 200 mA, which is included in the scope of delivery for this version. A replacement battery during a running series is not provided for in this series. Charging takes place via the power supply unit between two work steps.

Operating modes for multichannel applications

With frequent changes of medium within the same series, revPIP helps to dispense highly viscous or foaming reagents column by column without bubble formation.

Autoclave reprocessing

With the electronic version, unlike the mechanical Transferpette S -8, only the unscrewable pipette shaft is autoclaved, 121 °C, 2 bar, at least 15 minutes according to DIN EN 285. After that, a reference run on the handle section is required for all 8 channels together before operation continues.

Digital adjustment

The pipette can be adjusted digitally via the menu control, and Easy Calibration works in all modes except GEL. An adjustment made in PIP mode automatically also applies to PIPmix and revPIP, whereas a DISP adjustment remains limited to DISP. With 8 channels, an adjustment made once affects all channels equally, and readjusting individual channels is not provided for.

Typical applications

  • Dilution series column by column across eight channels
  • Portioning DNA and RNA samples in parallel
  • Adding reagents column by column in microtitre plates

Selection guide

The neighbouring variant 0.5-10 µl (MPN 705399) lies below, and the 5-100 µl version (MPN 705403) lies above. Number of channels, operating modes and adjustment procedure remain identical when changing between the volume steps of the same series. Only the test range, tip size and order number change.

Place within the Transferpette electronic family

In addition to the 8-channel version, there is the electronic multichannel pipette Transferpette -12 electronic with 12 channels in the same model system, and the electronic single-channel series Transferpette electronic is available for individual samples. Operating modes, adjustment procedure, battery and power supply are uniform across the whole electronic series, so laboratories with several channel numbers do not have to get used to different operating logic. Anyone who also wants to pipette mechanically will find the same number of channels in the Transferpette S -8 series.

Accuracy across the test range

Testing is carried out at three test volumes (100 %, 50 % and 10 % of the nominal volume). The systematic error is at 20 µl: ± 1.0 % / 10 µl: ± 2.0 % / 2 µl: ± 8.0 %, and the random error (coefficient of variation) at 20 µl: ± 0.5 % / 10 µl: ± 1.0 % / 2 µl: ± 3.0 %. Each of the 8 channels is subject to the same tolerance. As with any air cushion pipette, the relative deviation increases towards the lower test point and is tightest at the upper test point.

Technical details

FeatureValue
Number of channels8
Versionelectronic, multichannel
Volume range1-20 µl
Pipette tip0.5-20 µl
Systematic error (100/50/10 %)20 µl: ± 1.0 % / 10 µl: ± 2.0 % / 2 µl: ± 8.0 %
Random error CV (100/50/10 %)20 µl: ± 0.5 % / 10 µl: ± 1.0 % / 2 µl: ± 3.0 %
Operating modesPIP, PIPmix, revPIP, GEL, DISP
Autoclavabilitypipette shaft only, 121 °C, 2 bar, min. 15 min according to DIN EN 285, reference run afterwards
BatteryNiMH, 3x AAA, 3.6 V, 700 mAh
Power supply6.5 V DC, 200 mA, input 100-240 V/50-60 Hz, included in the scope of delivery
AdjustmentEasy Calibration digital via menu control, in every mode except GEL
DE-M markingunder German measurement and verification law
Manufacturer part no. (MPN)705400

Scope of delivery

  • Transferpette -8 electronic, volume range 1-20 µl, MPN 705400
  • Quality certificate
  • Battery (fitted at the factory)
  • Power supply for Europe
  • Silicone grease

Pipette tips are not included in the scope of delivery and are available separately.

Procurement and service

Accessories such as pipette tips, replacement batteries, power supplies or calibration services are put together on request by our procurement service. Details on our service page.

Number of channels: 8-channel
Operation: electronic
Volume range: 1 µl - 20 µl
Volume setting: variable