Skip to main content Skip to search Skip to main navigation

Brand Transferpette electronic Single-Channel Pipette, 10-200 µl, European Power Supply, DE-M, 705303

ⓘ  This product description has been translated from German using AI. Despite careful review, we do not guarantee that the translation is error-free; in case of doubt, the German version shall prevail.

Product information "Brand Transferpette electronic Single-Channel Pipette, 10-200 µl, European Power Supply, DE-M, 705303"

Standard pipetting in the mid microlitre range shapes the everyday work of many laboratories, from ELISA washing steps to dilution series. The Transferpette electronic with 10-200 µl covers this range electronically, including a mains adapter for Europe.

Display and operation

The display with rocker and selector wheel controls volume, mode and dispensing speed. Instead of the thumb, the motorised drive performs the actual piston stroke, and the trigger is sufficient for operation. Especially during long series of pipetting in the 10-200 µl range, this noticeably relieves the hand compared with a purely mechanical pipette, particularly over an entire work shift.

Battery and mains adapter in everyday use

A full charge of the built-in NiMH battery made of three individual AAA cells (3.6 V, 700 mAh) lasts for around 8 hours of continuous pipetting or more than 4000 pipetting cycles. A mains adapter for Europe is included in the scope of delivery (output voltage 6.5 V DC, 200 mA, order no. 705350), via which the battery is recharged directly in the charging stand.

Operating modes for different applications

Five operating modes are available: PIP for standard pipetting with aspiration and dispensing, PIPmix for continuous aspirating and dispensing when mixing, revPIP for reverse pipetting with highly viscous, high-vapour-pressure or foaming media, GEL for slow variable dispensing when loading electrophoresis gels, and DISP for the metered dispensing of a previously aspirated volume in several partial steps. The currently active mode is shown on the display and can be changed via the selector wheel without putting the pipette down.

Digital adjustment

Digital adjustment via Easy Calibration is available in every mode except GEL. If adjusted in PIP mode, PIPmix and revPIP take over this setting automatically, whereas an adjustment in DISP mode only applies there.

Autoclave reprocessing

Unlike the mechanical Transferpette S, in the electronic series the complete instrument is not autoclaved, but only the screw-off pipette shaft, 121 °C, 2 bar, at least 15 minutes according to DIN EN 285. After autoclaving, a reference run on the handle is required before pipetting again.

Selection guide

Below it is the neighbouring variant 2-20 µl (MPN 705300 or 705340), above it the version with 50-1,000 µl (MPN 705306 or 705346) for larger volumes. With the mains adapter included in the scope of delivery, the pipette is ready for use immediately. The operating logic remains the same across the whole electronic series; when changing the volume range, only test values, tip size and the associated order number change.

Typical applications

  • ELISA washing steps and standard assays in routine operation
  • Dilution series and sample preparation dosed electronically
  • Serial dispensing in DISP mode across several vessels

Accuracy over the test range

Testing is carried out at three test volumes (100 %, 50 % and 10 % of the nominal volume). The systematic error is 200 µl: ± 0.8 % / 100 µl: ± 1.2 % / 20 µl: ± 4.0 %, the random error (coefficient of variation) 200 µl: ± 0.2 % / 100 µl: ± 0.3 % / 20 µl: ± 0.6 %. For critical batches it follows that volumes close to the upper test point are reproducible most reliably, whereas smaller partial volumes deserve additional attention, especially with sensitive reagents.

Technical details

FeatureValue
Versionelectronic, single-channel
Volume range10-200 µl
Pipette tip2-200 µl
Systematic error (100/50/10 %)200 µl: ± 0.8 % / 100 µl: ± 1.2 % / 20 µl: ± 4.0 %
Random error CV (100/50/10 %)200 µl: ± 0.2 % / 100 µl: ± 0.3 % / 20 µl: ± 0.6 %
Operating modesPIP, PIPmix, revPIP, DISP, GEL
Autoclavabilitypipette shaft only, 121 °C, 2 bar, min. 15 min according to DIN EN 285, then reference run
BatteryNiMH, 3x AAA, 3.6 V, 700 mAh, approx. 8 h of continuous pipetting or over 4000 cycles
Mains adapter6.5 V DC, 200 mA, order no. 705350, included in the scope of delivery
AdjustmentEasy Calibration digital via menu control, in every mode except GEL
DE-M markingunder German measurement and verification law
Manufacturer art. no. (MPN)705303
EAN4033378418128

Scope of delivery

  • Transferpette electronic, volume range 10-200 µl, MPN 705303
  • Quality certificate
  • Battery (installed at the factory)
  • Mains adapter for Europe
  • Silicone grease

Pipette tips are not included in the scope of delivery and are available separately.

Procurement and service

Accessories such as pipette tips, replacement batteries, mains adapters or calibration services are put together by our procurement service on request. Details on our service page.

Number of channels: 1-channel
Operation: electronic
Volume range: 10 µl - 200 µl
Volume setting: variable