Brand Transferpette electronic Single-Channel Pipette, 250-5,000 µl, without Power Supply, DE-M, 705347
€340.00
- Electronic single-channel pipette, volume range 250-5,000 µl
- Operating modes: PIP, PIPmix, revPIP, DISP (without GEL)
- NiMH battery, 3x AAA, 3.6 V, 700 mAh, approx. 8 h continuous pipetting
- Without power supply, available to order separately
- Only the pipette shaft is autoclavable, 121 °C, 2 bar, reference run afterwards
- Matching tip: 500-5,000 µl
- DE-M marking under German measurement and verification law
ⓘ This product description has been translated from German using AI. Despite careful review, we do not guarantee that the translation is error-free; in case of doubt, the German version shall prevail.
Product information "Brand Transferpette electronic Single-Channel Pipette, 250-5,000 µl, without Power Supply, DE-M, 705347"
Larger liquid quantities of up to 5 ml, for example when preparing culture media in cell culture, call for a pipette with a correspondingly large nominal volume. This 250-5,000 µl version is supplied without a power supply.
Battery and power supply in everyday use
Three cylindrical AAA single cells (3.6 V, 700 mAh, NiMH) form the built-in battery, which lasts around 8 hours of continuous pipetting or over 4000 cycles when fully charged. This version is supplied without a power supply. The battery is fitted at the factory, and a suitable power supply for Europe (order no. 705350) can be ordered separately.
Digital adjustment
Easy Calibration is carried out digitally via the menu control and is possible in every mode except GEL. An adjustment in PIP mode is automatically transferred to PIPmix and revPIP, whereas an adjustment in DISP mode applies exclusively to DISP and is not transferred when the mode is changed.
Display and operation
The display with rocker switch and selector wheel controls volume, mode and dispensing speed. Instead of the thumb, the motorised drive carries out the actual piston stroke, and the trigger is all that is needed for operation. Especially during long series of pipetting in the 250-5,000 µl range, this noticeably relieves the hand compared with a purely mechanical pipette, particularly over a whole working shift.
Operating modes for different applications
Four of the five operating modes are available: PIP for standard pipetting, PIPmix for repeated aspirating and dispensing while mixing, revPIP for reverse pipetting of highly viscous, high-vapour-pressure or foaming media, and DISP for dispensing a previously aspirated volume in several partial steps. According to the operating instructions, the GEL mode for loading electrophoresis gels is not available for this volume range. The active mode is shown on the display and can be changed with the selector wheel without putting the pipette down.
Autoclave reprocessing
Unlike the mechanical Transferpette S, the electronic series is not autoclaved as a complete device. Only the unscrewable pipette shaft is autoclaved, 121 °C, 2 bar, at least 15 minutes according to DIN EN 285. After autoclaving, a reference run on the handle section is required before pipetting again.
Selection guide
The neighbouring variant 50-1,000 µl (MPN 705306 or 705346) lies below. For volumes beyond 5 ml, the mechanical Transferpette S variable is available with larger models. If you already own a suitable power supply, this version without a power supply saves unnecessary accessories. The operating logic is the same across the whole electronic series. When changing the volume range, only the test values, tip size and the corresponding order number change.
Typical applications
- Preparing culture media in cell culture (assuming an existing power supply)
- Decanting larger liquid quantities of up to 5 ml
- Filling larger reaction vessels with DISP serial dispensing
Accuracy across the test range
Testing is carried out at three test volumes (100 %, 50 % and 10 % of the nominal volume). The systematic error is at 5,000 µl: ± 0.6 % / 2,500 µl: ± 1.0 % / 500 µl: ± 3.0 %, and the random error (coefficient of variation) at 5,000 µl: ± 0.2 % / 2,500 µl: ± 0.3 % / 500 µl: ± 0.6 %. In practice, anyone dosing close to the nominal volume benefits from the tightest reproducibility. Smaller partial volumes should be cross-checked for sensitive dilution series before a larger series is started.
Technical details
| Feature | Value |
| Version | electronic, single-channel |
| Volume range | 250-5,000 µl |
| Pipette tip | 500-5,000 µl |
| Systematic error (100/50/10 %) | 5,000 µl: ± 0.6 % / 2,500 µl: ± 1.0 % / 500 µl: ± 3.0 % |
| Random error CV (100/50/10 %) | 5,000 µl: ± 0.2 % / 2,500 µl: ± 0.3 % / 500 µl: ± 0.6 % |
| Operating modes | PIP, PIPmix, revPIP, DISP (GEL not available for this volume range) |
| Autoclavability | pipette shaft only, 121 °C, 2 bar, min. 15 min according to DIN EN 285, reference run afterwards |
| Battery | NiMH, 3x AAA, 3.6 V, 700 mAh, approx. 8 h continuous pipetting or over 4000 cycles |
| Power supply | 6.5 V DC, 200 mA, order no. 705350, available separately |
| Adjustment | Easy Calibration digital via menu control, in every mode except GEL |
| DE-M marking | under German measurement and verification law |
| Manufacturer part no. (MPN) | 705347 |
Scope of delivery
- Transferpette electronic, volume range 250-5,000 µl, MPN 705347
- Quality certificate
- Battery (fitted at the factory)
- Silicone grease
Pipette tips are not included in the scope of delivery and are available separately.
Procurement and service
Accessories such as pipette tips, replacement batteries, power supplies or calibration services are put together on request by our procurement service. Details on our service page.
| Number of channels: | 1-channel |
|---|---|
| Operation: | electronic |
| Volume range: | 250 µl - 5000 µl |
| Volume setting: | variable |