Brand Transferpette electronic single-channel pipette, 50-1,000 µl, European power supply, DE-M, 705306
€358.00
- Electronic single-channel pipette, volume range 50-1,000 µl
- Operating modes: PIP, PIPmix, revPIP, DISP (without GEL)
- NiMH battery, 3x AAA, 3.6 V, 700 mAh, approx. 8 h continuous pipetting
- European power supply included in scope of delivery
- Only pipette shaft autoclavable, 121 °C, 2 bar, then reference run
- Suitable tip: 50-1,000 µl
- DE-M marking under German measurement and verification law
ⓘ This product description has been translated from German using AI. Despite careful review, we do not guarantee that the translation is error-free; in case of doubt, the German version shall prevail.
Product information "Brand Transferpette electronic single-channel pipette, 50-1,000 µl, European power supply, DE-M, 705306"
When dividing larger sample volumes or preparing buffers, volumes in the three-digit microlitre to low millilitre range regularly occur. The Transferpette electronic with 50-1,000 µl handles this electronically, power supply for Europe included in scope of delivery.
Operating modes for different applications
Four of the five operating modes are available: PIP for standard pipetting, PIPmix for continuous aspiration and dispensing during mixing, revPIP for reverse pipetting with highly viscous, high-vapour-pressure or foaming media and DISP for the stepped dispensing of a previously aspirated volume in multiple steps. According to the operating instructions, the GEL mode for loading electrophoresis gels is not available in this volume range. The currently active mode is shown on the display and can be changed using the selection thumbwheel without putting the pipette down.
Battery and power supply in everyday use
The NiMH battery consists of three individual AAA cells with 3.6 V and 700 mAh and, when fully charged, provides approximately 8 hours of continuous operation or over 4000 pipetting cycles. Included in the scope of delivery is a power supply for Europe (output voltage 6.5 V DC, 200 mA, order no. 705350), with which the battery is recharged directly at the charging stand.
Display and operation
The rocker key and selection thumbwheel on the display are used to adjust the volume, mode and dispensing speed. The motorised drive performs the piston stroke, the thumb merely operates the trigger. Especially during long serial pipetting runs in the 50-1,000 µl range, this noticeably relieves hand strain compared to a purely mechanical pipette, particularly across an entire work shift.
Digital adjustment
In all modes except GEL, digital adjustment can be performed via the menu control (Easy Calibration). A PIP adjustment is automatically transferred to PIPmix and revPIP, while a DISP adjustment applies separately only to the DISP mode.
Reprocessing in the autoclave
Unlike the mechanical Transferpette S, the entire device is not autoclaved in the electronic series, but only the unscrewable pipette shaft, 121 °C, 2 bar, at least 15 minutes according to DIN EN 285. After autoclaving, a reference run on the handle is required before pipetting again.
Selection guide
Below this is the neighbouring variant 10-200 µl (MPN 705303 or 705343), above it the version with 250-5,000 µl (MPN 705307 or 705347) for even larger volumes. With the power supply included in the scope of delivery, the pipette is ready for immediate use. The operating logic remains identical across the entire electronic series; when changing the volume range, only test values, tip size and the associated order number are changed.
Typical applications
- Sample splitting and buffer preparations in the millilitre range
- Transferring larger sample volumes between vessels
- Preparation of stock solutions with DISP serial dispensing
Accuracy across the test range
Testing is carried out at three test volumes (100 %, 50 % and 10 % of nominal volume). The systematic error is at 1,000 µl: ± 0.6 % / 500 µl: ± 1.0 % / 100 µl: ± 3.0 %, the random error (coefficient of variation) at 1,000 µl: ± 0.2 % / 500 µl: ± 0.3 % / 100 µl: ± 0.6 %. In practice, this means that the closer to the nominal volume the work is performed, the tighter the reproducibility; for smaller partial volumes, taking a spot check before the actual series is worthwhile for critical series.
Technical details
| Feature | Value |
| Version | electronic, single-channel |
| Volume range | 50-1,000 µl |
| Pipette tip | 50-1,000 µl |
| Systematic error (100/50/10 %) | 1,000 µl: ± 0.6 % / 500 µl: ± 1.0 % / 100 µl: ± 3.0 % |
| Random error CV (100/50/10 %) | 1,000 µl: ± 0.2 % / 500 µl: ± 0.3 % / 100 µl: ± 0.6 % |
| Operating modes | PIP, PIPmix, revPIP, DISP (GEL not available for this volume range) |
| Autoclavability | only pipette shaft, 121 °C, 2 bar, min. 15 min according to DIN EN 285, then reference run |
| Battery | NiMH, 3x AAA, 3.6 V, 700 mAh, approx. 8 h continuous pipetting or over 4000 cycles |
| Power supply | 6.5 V DC, 200 mA, order no. 705350, included in scope of delivery |
| Adjustment | Easy Calibration digital via menu control, in every mode except GEL |
| DE-M marking | under German measurement and verification law |
| Manufacturer part no. (MPN) | 705306 |
Scope of delivery
- Transferpette electronic, volume range 50-1,000 µl, MPN 705306
- Quality certificate
- Battery (factory-fitted)
- Power supply for Europe
- Silicone grease
Pipette tips are not included in the scope of delivery and are available separately.
Procurement and service
Accessories such as pipette tips, replacement batteries, power supplies or calibration services are compiled on request by our procurement service. Details on our service page.
| Number of channels: | 1-channel |
|---|---|
| Operation: | electronic |
| Volume range: | 50 µl - 1000 µl |
| Volume setting: | variable |