Skip to main content Skip to search Skip to main navigation

Brand Transferpette S -8 Multichannel Pipette, 8-Channel, 20-200 µl, DE-M, 705910

ⓘ  This product description has been translated from German using AI. Despite careful review, we do not guarantee that the translation is error-free; in case of doubt, the German version shall prevail.

Product information "Brand Transferpette S -8 Multichannel Pipette, 8-Channel, 20-200 µl, DE-M, 705910"

Many assay protocols work with substrate or wash volumes in the range of 20 to 200 µl. The Transferpette S -8 in this volume step covers exactly this window column by column across eight channels.

The 20-200 µl range is covered in model M8-200 by 8 individually manufactured and tested pistons, and the increment is 0.2 µl. Tips in the 2-200 µl range sit firmly on all channels and can be removed without contact via the ejection mechanism. Despite 8 pistons guided in parallel, the effort remains close to that of a single-channel pipette.

Adjustment without tools

Instead of an external tool, the adjustment slider under the label window is enough to correct the volume. A volume offset supplements this adjustment for media of different density or for different site altitudes.

Reprocessing in the autoclave

The Transferpette S -8 is autoclavable as a whole, 121 °C, 2 bar, at least 15 minutes holding time according to DIN EN 285, without dismantling before the sterilisation cycle. On the electronic sister line, the Transferpette -8 electronic, only the unscrewable pipette shaft can be autoclaved, which is a significant difference between the mechanical and the electronic version.

Ergonomics in series operation

For continuous use on the microtitre plate, the pipette body of the Transferpette S -8 is designed so that 8 pistons guided in parallel do not overtax the span of the hand. The finger stroke remains even across 20-200 µl. The grip section and tip cone are kept neutral, and the number of channels is clearly legible on the housing.

Place within the Transferpette S family

In addition to the 8-channel version, the Transferpette S family also includes the variant with 12 channels (Transferpette S -12) and the single-channel line Fix and Variable for individual samples. Since the adjustment method and autoclavability are the same across the line, operation remains unchanged when switching between channel counts.

Selection guide

Below it is the neighbouring variant with 10-100 µl (MPN 705908), above it the version with 30-300 µl (MPN 705912) for even larger volumes. Channel count, model and adjustment method stay identical when switching between volume steps of the same line, only the test range and the corresponding tip size change. If you are unsure which range suits your protocol, base the choice on the volume you pipette most often within the series and pick the range in which this volume lies as close as possible to the upper test point.

Typical applications

  • ELISA substrate and wash steps column by column
  • larger assay volumes in parallel across eight wells
  • dilution series in the medium microlitre range column by column

Accuracy across the test range

Testing is carried out at three test volumes (100 %, 50 % and 10 % of the nominal volume 20-200 µl). The systematic measurement error is given as 200 µl: ± 0.8 % / 100 µl: ± 1.4 % / 20 µl: ± 4 %, the random measurement error (coefficient of variation) as 200 µl: ± 0.3 % / 100 µl: ± 0.6 % / 20 µl: ± 1.5 %. As with any air cushion pipette, the relative deviation grows towards the lower test point and is tightest at the upper test point. Anyone working close to the nominal volume benefits from the tightest reproducibility, while smaller partial volumes within the range should additionally be checked in sensitive dilution series.

Technical details

FeatureValue
Number of channels8
Volume range20-200 µl
Increment0.2 µl
ModelM8-200
Pipette tip2-200 µl
Systematic measurement error (100/50/10 %)200 µl: ± 0.8 % / 100 µl: ± 1.4 % / 20 µl: ± 4 %
Random measurement error CV (100/50/10 %)200 µl: ± 0.3 % / 100 µl: ± 0.6 % / 20 µl: ± 1.5 %
Autoclavabilitycomplete, 121 °C, 2 bar, at least 15 min according to DIN EN 285
AdjustmentEasy Calibration, adjustment slider under the label window, volume offset
Manufacturer part no. (MPN)705910
EAN4033378493934

Scope of delivery

  • Transferpette S -8, model M8-200, volume range 20-200 µl, MPN 705910
  • Quality certificate
  • Silicone grease

Pipette tips are not included in the scope of delivery and are available separately. When ordering several channel counts and volume ranges, a combined collective order is recommended to bundle delivery times and shipping costs.

Procurement and service

Accessories such as pipette tips, spare parts or calibration services are assembled by our procurement service on request. Details on our service page.

Number of channels: 8-channel
Operation: mechanical
Volume range: 20 µl - 200 µl
Volume setting: variable