Brand Transferpettor Positive Displacement Pipette, Fixed, 25 µl, with Glass Capillary, 701864

ⓘ  This product description has been translated from German using AI. Despite careful review, we do not guarantee that the translation is error-free; in case of doubt, the German version shall prevail.

Product information "Brand Transferpettor Positive Displacement Pipette, Fixed, 25 µl, with Glass Capillary, 701864"

Concentrated surfactant standards foam up quickly in an air-cushion pipette; a fixed volume with direct displacement avoids this problem from the outset. The Brand Transferpettor positive displacement pipette, fixed at 25 µl, operates without an air cushion and with a fixed volume. When preparing surfactant standards, residual wetting in the glass capillary therefore remains low, regardless of how much the medium foams.

Direct displacement principle without an air cushion

Like the micro models in the Digital range, the 25 µl fixed-volume version also features a glass capillary in which the piston moves without an air cushion. According to Brand, media with a density of up to 3.8 g/cm³ and a vapour pressure of up to 500 mbar are approved; the operating temperature range is 15 to 40 °C. According to the manufacturer, the kinematic viscosity limit for the nominal volume of 25 µl is up to 50,000 mm²/s (nominal volume 25 µl). As the volume cannot be adjusted, the Fix series does not specify a graduation increment.

The colour code for this 25 µl version is 2x white; the accuracy is 2.4 % and the coefficient of variation is 0.4 %.

Typical applications

Typical applications for the 25 µl version include the following:

  • The preparation of concentrated surfactant standards
  • Handling highly volatile reagents, where an air-cushion pipette produces inaccurate results due to evaporation within the air cushion
  • Working with foaming or highly viscous media such as surfactant solutions, where an air-cushion pipette traps air bubbles

Selection guide for the Fix range

For the nominal volume of 25 µl, the manufacturer specifies a kinematic viscosity limit of up to 50,000 mm²/s, which is significantly higher than for the Fix models up to 10 µl. The design remains a glass capillary, as with the entire Fix range. Anyone requiring additional variable volumes between 10 and 50 µl can supplement this Fix pipette with the 10-50 µl Digital version.

Fix or Digital

If the volume remains the same throughout the entire workflow, the simpler, more affordable Fix version is preferable, as is the case here with a 25 µl capacity. If the volume varies from sample to sample, for example, in dilution series, a Digital version from the same product range is the more suitable choice, as it can be adjusted to a new value without having to change the pipette.

Maintenance and accessories

For the fixed-volume range, we offer the same range of accessories as for the micro version of the Digital range: station, glass capillaries, seals, Caps+Seal combo pack, repair kit, micro piston rod, clamping ring and attachment block. The repair kit covers the replacement of the piston rod and seal.

For the 25 µl version with colour code 2x white, this range of accessories remains unchanged, regardless of the specific fixed volume.

Frequently Asked Questions

Can the volume of the 25 µl version be adjusted?
No, the Fix series has a fixed volume without an adjustment wheel; due to its design, adjustment is not provided for.

How accurate is this volume setting?
The accuracy is 2.4 %, with a coefficient of variation of 0.4 %, according to the manufacturer’s test data for the 25 µl version with colour code 2x white.

Which media are compatible with the Fix series?
For the entire Fix series with glass capillaries, Brand specifies a maximum density of 3.8 g/cm³ and a maximum vapour pressure of 500 mbar, at an operating temperature of 15 to 40 °C.

Which accessories are compatible with the 25 µl version?
The station, glass capillaries, seals, Caps+Seal combo pack, repair kit, micro piston rod, clamping ring and attachment block, identical to those for the micro version of the Digital range.

What is the 25 µl version typically used for?
For preparing concentrated surfactant standards, as well as for surfactant solutions of a constant volume in general.

Technical details

LineTransferpettor Fix
DesignGlass capillary
Volume25 µl (fixed volume)
Accuracy R2.4 %
Coefficient of variation CV0.4 %
Colour code2x white
Max. vapour pressure500 mbar
Max. density3.8 g/cm³
Operating temperature15 to 40 °C
Max. kinematic viscosity50,000 mm²/s (nominal volume 25 µl)

Scope of delivery

  • Transferpettor positive displacement pipette, 25 µl

Individual requirements and procurement service

If, in addition to the 25 µl version, you require a different volume size from the Transferpettor-Fix range, suitable accessories or a spare part, our procurement service will be happy to assist.

Number of channels: 1-channel
Operation: mechanical
Volume range: 25 µl (fixed)