Skip to main content Skip to search Skip to main navigation

Eppendorf Reference 2 single-channel pipette set, pack of 3, Option 2, 2–20 µL / 20–200 µL / 100–1,000 µL, 4924000916

ⓘ  This product description has been translated from German using AI. Despite careful review, we do not guarantee that the translation is error-free; in case of doubt, the German version shall prevail.

Product information "Eppendorf Reference 2 single-channel pipette set, pack of 3, Option 2, 2–20 µL / 20–200 µL / 100–1,000 µL, 4924000916"

Two pipettes, one tip: in the Option 2 set of the Eppendorf Reference 2, the pipettes for 2 to 20 µL and 20 to 200 µL both have the colour code yellow and take epT.I.P.S. 2 to 200 µL. The third, for 100 to 1,000 µL, has the colour code blue for epT.I.P.S. 50 to 1,000 µL. All three are mechanical single-channel pipettes with adjustable volume.

Yellow refers to the tip, not the volume

The 2 to 20 µL pipette is also available from Eppendorf in light grey, then with epT.I.P.S. 0.5 to 20 µL (length L). This set contains the yellow version. As a result, one supply of tips is enough for two of the three pipettes; only the 100 to 1,000 µL pipette needs a second tip size.

One-button operation

According to Eppendorf, volume selection, pipetting and tip ejection are carried out with a single control element. A magnifier window above the four-digit display makes the volume easy to read, and a volume lock holds it in place. Haptic feedback is intended to prevent accidental tip ejection according to Eppendorf; the data sheet also mentions active aerosol reduction. The pipettes work with an air cushion.

Typical applications

  • Preparing samples and reagents in the 2 to 20 µL range
  • Dispensing aliquots of 20 to 200 µL from stock solutions
  • Transferring buffers and media up to 1,000 µL
  • Serial dilutions covering several orders of magnitude

Autoclaving

All three pipettes may go into the autoclave completely at 121 °C and 1 bar for 20 minutes, without additional disinfectants or sodium hypochlorite. Filter sleeve and protective filter, where present, are autoclaved separately. The pipettes are UV-resistant at 254 nm.

Technical details

FeatureValue
SeriesEppendorf Reference 2, mechanical
Eppendorf designationOption 2
TypeSet of 3 single-channel pipettes, adjustable volume, mechanical
Pipette 12 to 20 µL, colour code yellow, epT.I.P.S. 2 to 200 µL
Pipette 220 to 200 µL, colour code yellow, epT.I.P.S. 2 to 200 µL
Pipette 3100 to 1,000 µL, colour code blue, epT.I.P.S. 50 to 1,000 µL
Weight of pipette 100 to 1,000 µL97 g according to the Eppendorf data sheet
Autoclavablefully, 121 °C, 1 bar, 20 minutes
StandardConformity with ISO 8655 according to Eppendorf
ManufacturerEppendorf
Manufacturer number4924000916

Measurement deviations

Eppendorf specifies three or four test volumes per pipette. According to Eppendorf, the values apply only with original epT.I.P.S. They are determined gravimetrically according to ISO 8655, with 10 determinations per volume and water according to ISO 3696. The systematic error describes how far the mean value deviates from the set volume, the random error the scatter of the individual values.

PipetteTest volumeSystematic errorRandom error
2 to 20 µL2 µL±5 % / ±0.1 µL±1.5 % / ±0.03 µL
10 µL±1.2% / ±0.12 µL±0.6 % / ±0.06 µL
20 µL±1 % / ±0.2 µL±0.3 % / ±0.06 µL
20 to 200 µL20 µL±2.5 % / ±0.5 µL±0.7 % / ±0.14 µL
100 µL±1 % / ±1 µL±0.3 % / ±0.3 µL
200 µL±0.6 % / ±1.2 µL±0.2 % / ±0.4 µL
100 to 1,000 µL100 µL±3 % / ±3 µL±0.6% / ±0.6 µL
500 µL±1 % / ±5 µL±0.2 % / ±1 µL
1 mL±0.6% / ±6 µL±0.2 % / ±2 µL

Scope of delivery

  • 3 single-channel pipettes (see table)
  • 3 epT.I.P.S. Box 2.0, each containing 96 tips
  • Ballpoint pen

Accessories and procurement

Pipette Carousel 2 and Pipette Stand 2 from Eppendorf are also available. The procurement service arranges further epT.I.P.S. on request.

Colour: Yellow
Number of channels: 1-channel
Operation: mechanical
Volume setting: variable