Eppendorf Reference 2 single-channel pipette set, pack of 3, Option 2, 2–20 µL / 20–200 µL / 100–1,000 µL, 4924000916
- Set of 3 mechanical single-channel pipettes, Eppendorf Option 2
- 2 to 20 µL, 20 to 200 µL and 100 to 1,000 µL, one-button operation
- Colour code yellow, yellow and blue: two pipettes take epT.I.P.S. 2 to 200 µL
- 3 epT.I.P.S. Box 2.0 and a ballpoint pen included in the scope of delivery
- Fully autoclavable at 121 °C, UV-resistant at 254 nm
- Conforms to ISO 8655 according to Eppendorf, measurement deviation specified for each pipette
ⓘ This product description has been translated from German using AI. Despite careful review, we do not guarantee that the translation is error-free; in case of doubt, the German version shall prevail.
Product information "Eppendorf Reference 2 single-channel pipette set, pack of 3, Option 2, 2–20 µL / 20–200 µL / 100–1,000 µL, 4924000916"
Two pipettes, one tip: in the Option 2 set of the Eppendorf Reference 2, the pipettes for 2 to 20 µL and 20 to 200 µL both have the colour code yellow and take epT.I.P.S. 2 to 200 µL. The third, for 100 to 1,000 µL, has the colour code blue for epT.I.P.S. 50 to 1,000 µL. All three are mechanical single-channel pipettes with adjustable volume.
Yellow refers to the tip, not the volume
The 2 to 20 µL pipette is also available from Eppendorf in light grey, then with epT.I.P.S. 0.5 to 20 µL (length L). This set contains the yellow version. As a result, one supply of tips is enough for two of the three pipettes; only the 100 to 1,000 µL pipette needs a second tip size.
One-button operation
According to Eppendorf, volume selection, pipetting and tip ejection are carried out with a single control element. A magnifier window above the four-digit display makes the volume easy to read, and a volume lock holds it in place. Haptic feedback is intended to prevent accidental tip ejection according to Eppendorf; the data sheet also mentions active aerosol reduction. The pipettes work with an air cushion.
Typical applications
- Preparing samples and reagents in the 2 to 20 µL range
- Dispensing aliquots of 20 to 200 µL from stock solutions
- Transferring buffers and media up to 1,000 µL
- Serial dilutions covering several orders of magnitude
Autoclaving
All three pipettes may go into the autoclave completely at 121 °C and 1 bar for 20 minutes, without additional disinfectants or sodium hypochlorite. Filter sleeve and protective filter, where present, are autoclaved separately. The pipettes are UV-resistant at 254 nm.
Technical details
| Feature | Value |
|---|---|
| Series | Eppendorf Reference 2, mechanical |
| Eppendorf designation | Option 2 |
| Type | Set of 3 single-channel pipettes, adjustable volume, mechanical |
| Pipette 1 | 2 to 20 µL, colour code yellow, epT.I.P.S. 2 to 200 µL |
| Pipette 2 | 20 to 200 µL, colour code yellow, epT.I.P.S. 2 to 200 µL |
| Pipette 3 | 100 to 1,000 µL, colour code blue, epT.I.P.S. 50 to 1,000 µL |
| Weight of pipette 100 to 1,000 µL | 97 g according to the Eppendorf data sheet |
| Autoclavable | fully, 121 °C, 1 bar, 20 minutes |
| Standard | Conformity with ISO 8655 according to Eppendorf |
| Manufacturer | Eppendorf |
| Manufacturer number | 4924000916 |
Measurement deviations
Eppendorf specifies three or four test volumes per pipette. According to Eppendorf, the values apply only with original epT.I.P.S. They are determined gravimetrically according to ISO 8655, with 10 determinations per volume and water according to ISO 3696. The systematic error describes how far the mean value deviates from the set volume, the random error the scatter of the individual values.
| Pipette | Test volume | Systematic error | Random error |
|---|---|---|---|
| 2 to 20 µL | 2 µL | ±5 % / ±0.1 µL | ±1.5 % / ±0.03 µL |
| 10 µL | ±1.2% / ±0.12 µL | ±0.6 % / ±0.06 µL | |
| 20 µL | ±1 % / ±0.2 µL | ±0.3 % / ±0.06 µL | |
| 20 to 200 µL | 20 µL | ±2.5 % / ±0.5 µL | ±0.7 % / ±0.14 µL |
| 100 µL | ±1 % / ±1 µL | ±0.3 % / ±0.3 µL | |
| 200 µL | ±0.6 % / ±1.2 µL | ±0.2 % / ±0.4 µL | |
| 100 to 1,000 µL | 100 µL | ±3 % / ±3 µL | ±0.6% / ±0.6 µL |
| 500 µL | ±1 % / ±5 µL | ±0.2 % / ±1 µL | |
| 1 mL | ±0.6% / ±6 µL | ±0.2 % / ±2 µL |
Scope of delivery
- 3 single-channel pipettes (see table)
- 3 epT.I.P.S. Box 2.0, each containing 96 tips
- Ballpoint pen
Accessories and procurement
Pipette Carousel 2 and Pipette Stand 2 from Eppendorf are also available. The procurement service arranges further epT.I.P.S. on request.
| Colour: | Yellow |
|---|---|
| Number of channels: | 1-channel |
| Operation: | mechanical |
| Volume setting: | variable |