Eppendorf Research plus single-channel pipette, variable, 0.5–5 mL, purple, 3123000071
- Mechanical single-channel pipette, 0.5 to 5 mL, step size 0.005 mL
- Four-digit volume display, air cushion
- Colour code purple for epT.I.P.S. 0.1 to 5 mL
- Fully autoclavable at 121 °C, UV-resistant at 254 nm
- Temporary adjustment with adjustment seals, return to the factory adjustment without calibration
- Conforms to ISO 8655 according to Eppendorf, 3 test volumes with measurement deviation specified
ⓘ This product description has been translated from German using AI. Despite careful review, we do not guarantee that the translation is error-free; in case of doubt, the German version shall prevail.
Product information "Eppendorf Research plus single-channel pipette, variable, 0.5–5 mL, purple, 3123000071"
The Eppendorf Research plus for 0.5 to 5 mL is a mechanical single-channel pipette with adjustable volume. The volume is selected in steps of 0.005 mL and read on the four-digit volume display. The colour code purple belongs to the matching tip, namely epT.I.P.S. 0.1 to 5 mL.
Setting the volume
Volume selection, pipetting and tip ejection are handled by separate controls. The pipette works with an air cushion. The smallest adjustable step is 0.005 mL. Anyone who pipettes preparations in many different volumes manages with an adjustable-volume pipette and does not need a set of fixed-volume pipettes.
Tips and colour code
According to Eppendorf, the colour of the pipette corresponds to the colour of the matching epT.I.P.S. Here, this is the colour code purple for epT.I.P.S. 0.1 to 5 mL. The colour code depends on the tip and not on the pipette volume. The pipette is an open system, and tips from other manufacturers are possible. However, the measurement deviations apply only with original epT.I.P.S.
Adjusting
The pipette can be adjusted temporarily. 5 adjustment seals and an adjustment tool (hex key) are supplied for this purpose. Returning to the factory adjustment is possible without calibration.
Typical applications
- Transferring reagents and samples into reaction vessels
- Preparing dilution series with changing volumes
- Aliquoting liquids in the range 0.5 to 5 mL
- Mixing solutions by repeated aspiration and dispensing
- Methods in which the volume changes from preparation to preparation
When choosing, it helps to look at the volume that occurs most often in daily work. If it is close to the nominal volume, the relative measurement deviation is smaller, as the table below shows.
Autoclaving and resistance
The entire pipette may go into the autoclave at 121 °C and 1 bar for 20 minutes. Disinfectants and decontamination agents are wiped off beforehand. Filter sleeve and protective filter, where present on the model, are autoclaved separately. The pipette is UV-resistant at 254 nm. According to Eppendorf, chemical resistance is high, and special lower parts are available for aggressive chemicals. The pipette is operated at +5 °C to +40 °C and 10 to 95 % relative humidity. For storage, Eppendorf specifies -5 °C to +45 °C.
Technical details
| Feature | Value |
|---|---|
| Series | Eppendorf Research plus, mechanical |
| Type | Single-channel pipette, adjustable volume |
| Volume range | 0.5 to 5 mL |
| Step size | 0.005 mL |
| Display | four-digit volume display |
| Colour code | purple |
| Matching tip | epT.I.P.S. 0.1 to 5 mL |
| Autoclavable | fully, 121 °C, 1 bar, 20 minutes |
| UV resistance | 254 nm |
| Operating conditions | +5 °C to +40 °C, 10 to 95 % relative humidity |
| Storage | -5 °C to +45 °C |
| Standard | Conformity with ISO 8655 according to Eppendorf |
| Manufacturer | Eppendorf |
| Manufacturer number | 3123000071 |
Measurement deviations
For this pipette, Eppendorf specifies 3 test volumes. According to Eppendorf, the values apply only with original epT.I.P.S. Eppendorf determines them gravimetrically according to ISO 8655, with 10 determinations per volume and water according to ISO 3696. The systematic error describes how far the mean value deviates from the set volume. The random error describes the scatter of the individual values.
| Test volume | Systematic error | Random error |
|---|---|---|
| 500 µL | ±2.4 % / ±12 µL | ±0.6 % / ±3 µL |
| 2,500 µL | ±1.2 % / ±30 µL | ±0.25 % / ±6.25 µL |
| 5,000 µL | ±0.6 % / ±30 µL | ±0.15 % / ±7.5 µL |
Scope of delivery
- Pipette
- Operating instructions
- Certificate
- Adjustment tool (hex key)
- 5 adjustment seals for temporary adjustment
- Pipette grease
Depending on the model, further parts are included, for example an epT.I.P.S. Box 2.0 or a sample bag with epT.I.P.S., a locking ring for deactivating the spring loading, protective filters and a pipette wrench.
Accessories and procurement
The Pipette Carousel 2 and the Pipette Stand 2 from Eppendorf are also available. Further epT.I.P.S. and accessories are procured by the procurement service on request.
| Colour: | Violet |
|---|---|
| Number of channels: | 1-channel |
| Operation: | mechanical |
| Volume range: | 0.5 ml - 5 ml |
| Volume setting: | variable |