Skip to main content Skip to search Skip to main navigation

Kern ABP 100-4M analytical balance 120 g / 0.0001 g

Product information "Kern ABP 100-4M analytical balance 120 g / 0.0001 g"

Anyone who needs to read weighed-in portions of standard solutions, formulations or samples to 0.1 mg or 0.01 mg will find seven models in the Kern ABP. The analytical balances weigh up to 320 g, use a single-cell weighing system and have an OLED display, a glass draught shield with three sliding doors and an automatic internal adjustment. Three models read to 0.0001 g (ABP 100-4M, 200-4M and 300-4M). Four models read to 0.00001 g; Kern calls the single-range models ABP 100-5M and 200-5M semi-micro balances. The ABP 100-5DM and 200-5DM are multi-range balances and switch automatically between two weighing ranges as the load increases or decreases.

Short stabilisation time and automatic adjustment

According to Kern, the stabilisation time is 2 s for the models with 0.0001 g and 8 s for the single-range models with 0.00001 g. The multi-range models state both values. The internal automatic adjustment adjusts the balance when the temperature changes by 1 °C or more and additionally every 4 h. This has nothing to do with a calibration certificate, which is available separately (see below).

The minimum sample weight can be stored in the balance or calculated automatically. If less is weighed in, the balance issues a warning message.

Formulating, users and data output

When formulating, the balance numbers the ingredients and prints them with number and weight. A combined tare/print key saves steps. Up to 10 users can be set up with password, menu language and permissions; they can also be called up by barcode. Anyone who is not logged in works in guest mode. Weighing data go to a PC and printer via RS-232 or USB device, also automatically each time the balance comes to rest. A USB keyboard can be used at the USB host port for item numbers and formulations. According to Kern, the balance supports data integrity according to U.S. FDA 21 CFR Part 11, for example with weighing result, sample ID, user name and balance ID.

Functions and accessories in the weighing chamber

Besides weighing, there are piece counting (minimum unit weight 1 mg under laboratory conditions, 10 mg under standard conditions), percentage determination, density determination for liquids and solids, measuring with tolerance range and below-balance weighing. Units are g, mg and ct. The stainless steel weighing pan has a diameter of 91 mm. According to Kern, a multi-function weighing plate is included in the scope of delivery. It reduces the influence of air currents and holds overhanging samples, sample paper, PCR tubes or microcentrifuge tubes in place.

Typical applications

  • Pharmaceutical applications: weighing in formulation components with a numbered printout
  • Quality control: weight checks with tolerance range and report
  • Research and development: sample weighing in the milligram range
  • Chemical industry, and industry and production: preparing solutions, density determination

The models with 0.0001 g are sufficient when 0.1 mg is enough. Anyone who regularly weighs in portions below 10 mg will choose a model with 0.00001 g, where the minimum load is 0.001 g instead of 0.01 g. Anyone who switches between small and large portions is well served by the multi-range models: the ABP 100-5DM weighs up to 52 g with a readability of 0.00001 g and up to 120 g with a readability of 0.0001 g, the ABP 200-5DM up to 102 g with a readability of 0.00001 g and up to 220 g with a readability of 0.0001 g.

Models at a glance

For the multi-range models, the values are listed in the order of the weighing ranges.

ModelWeighing range (Max)Readability (d)Verification value (e)Minimum load (Min)ReproducibilityLinearityStabilisation time
ABP 100-4M120 g0.0001 g0.001 g0.01 g0.0001 g± 0.0002 g2 s
ABP 100-5DM52 g / 120 g0.00001 g / 0.0001 g0.001 g0.001 g0.00002 g / 0.0001 g± 0.00005 g / ± 0.0002 g8 s / 2 s
ABP 100-5M135 g0.00001 g0.001 g0.001 g0.00005 g± 0.0001 g8 s
ABP 200-4M220 g0.0001 g0.001 g0.01 g0.0001 g± 0.0002 g2 s
ABP 200-5DM102 g / 220 g0.00001 g / 0.0001 g0.001 g0.001 g0.00005 g / 0.0001 g± 0.0001 g / ± 0.0002 g8 s / 2 s
ABP 200-5M220 g0.00001 g0.001 g0.001 g0.000015 g / 0.00003 g / 0.00005 g± 0.0001 g8 s
ABP 300-4M320 g0.0001 g0.001 g0.01 g0.00015 g± 0.0003 g2 s

Technical details

These values apply to all models. Deviating values are given in the model table.

Weighing systemSingle cell
Accuracy classI
Weighing panStainless steel, Ø 91 mm
DisplayOLED, digit height 12 mm
Unitsg, mg, ct
InterfacesRS-232 (standard), USB device, USB host
HousingDie-cast aluminium, plastic
Housing dimensions (W x D x H)220 x 370 x 350 mm
Net weightapprox. 8 kg
Power supplyPlug-in mains adaptor, 100 V to 240 V AC, 50/60 Hz
Ambient temperature10 °C to 30 °C
Humidityup to 80 %
Conformity assessment under NAWI (2014/31/EU)optional, available ex works only, item no. 965-201
DAkkS calibration certificatecan be ordered separately, item no. 963-101
Warranty (manufacturer)36 months according to Kern, excluding wear parts

Conformity assessment and calibration

All models have the verification value e = 0.001 g and accuracy class I. A conformity assessment under the NAWI Directive (2014/31/EU) is optional (item no. 965-201). It can only be ordered ex works together with the balance; according to Kern, a subsequent conformity assessment is not possible. Kern needs the full address of the installation site for this. A DAkkS calibration certificate (item no. 963-101) can be ordered separately. The confirmation of manufacturer specification (969-517) is only available in conjunction with the DAkkS calibration certificate.

Scope of delivery

  • Kern ABP analytical balance with stainless steel weighing pan (Ø 91 mm) and glass draught shield
  • Multi-function weighing plate
  • Working protective cover
  • Plug-in mains adaptor with country adaptors for EU, AUS, UK, US and CH
  • ABP 200-5M only: Erlenmeyer flask holder

Warranty

According to Kern & Sohn, these devices come with a 3-year warranty. Kern grants 3 years on products with a list price of €500 or more and 2 years on products below that and on components; wear parts such as batteries and rechargeable batteries are excluded. Warranty repairs are carried out by Kern or by a Kern dealer.

Downloads

Regulatory information

As electrical devices, the balances fall under the ElektroG. The manufacturer Kern & Sohn GmbH is registered under the WEEE registration number DE 80821200. How to return old appliances is described on the page Environment and disposal.

Individual requirements and procurement service

Anti-vibration table (YPS-03), anti-vibration plates (YPS-04, YPS-05), dust cover (ABS-A08), draught shield (ABP-A02), ioniser (ABP-A01), density determination kit (YDB-03), thermal printers (YKC-01, YKH-01), RS232 Bluetooth adapter (YKI-02) and USB barcode scanner (YKX-A03) are also available. Enquiries are handled via our procurement service.

Weighing range (Max): 120 g
Readability (d): 0.0001 g
Verifiable: yes
Design: ABP 100-4M (120 g / 0,0001 g)
Power supply: Mains operation