Kern DIS 50-4 moisture analyser 50 g / 0.0001 g, up to 175 °C

ⓘ  This product description has been translated from German using AI. Despite careful review, we do not guarantee that the translation is error-free; in case of doubt, the German version shall prevail.

Product information "Kern DIS 50-4 moisture analyser 50 g / 0.0001 g, up to 175 °C"

The Kern DIS is available in four models, which differ in two respects: the weighing range and the temperature range. The DIS 50-4 weighs up to 50 g to an accuracy of 0.0001 g (0.1 mg), whilst the DIS 100-3 weighs up to 120 g to an accuracy of 0.001 g (1 mg). The two standard models heat from 30 °C to 175 °C, whilst the HT models, the DIS 50-4HT and DIS 100-3HT, heat up to 250 °C, in 1 °C increments in each case. The heat is provided by a 400 W halogen quartz glass heater.

Which model is suitable?

For small samples, the DIS 50-4 is the more accurate choice. For a 2 g initial weight, the data sheet specifies a reproducibility of 0.05 %; for the DIS 100-3, this figure is 0.1 %. The DIS 100-3, on the other hand, can weigh up to 120 g. The HT version is designed for samples that are to be dried at temperatures above 175 °C. According to Kern, the extended temperature range allows for adaptation to different material properties and reduces the measurement time for samples with stable temperatures.

Operation in three steps

The measurement process consists of three steps: taring, adding the sample, and starting. The backlit graphic LCD (digit height 15 mm) displays the drying curve in real time; the last measured value remains on display until the next measurement. The HT models also feature a configurable display: the result can be shown as either moisture content, dry weight, ATRO value or residual weight, and the digit display, contrast and backlighting are adjustable. Password protection prevents stored settings and data from being altered.

Heating profiles, memory and log

There are four heating profiles to choose from: standard drying, rapid drying, gentle drying and step drying. Up to 40 drying programmes can be stored. Drying ends automatically, by timer or manually.

The instrument stores up to 1,000 measurement results and up to 1,000 freely programmable sample names, which are printed on the measurement report. Date and time are included as standard. Data is transferred to a PC or printer via RS-232 or USB; the GLP/ISO report is an instrument function for printing, not an external certification. The weighing system, which features electromagnetic force compensation, is equipped with overload protection.

Installation

The instrument measures 211 x 342 x 187 mm (W x D x H) and weighs approx. 5.0 kg. It is designed for ambient temperatures of 15 °C to 30 °C and relative humidity of 25 % to 85 %. The power supply is built-in (230 V AC, 50 Hz); connection is via an IEC power cable. The balance is adjusted using an external weight.

Typical applications

  • Food processing and quality control
  • Pharmaceutical and chemical industries, as well as analytical and chemical laboratories
  • Cosmetics manufacturing and the plastics and polymers industry
  • Building materials and cement industries, as well as the ceramics and non-metallic materials industries
  • Agricultural product monitoring and quality assurance
  • Water and wastewater management (e.g. sewage treatment works) and energy generation from biomass

Models at a glance

ModelWeighing range (Max)Readability (d)LinearityStabilisation timeTemperature range (1 °C steps)Reproducibility at sample weight 10 g / 2 gUSP minimum sample weight (k = 2, U = 0.1 %)
DIS 50-450 g0.0001 g± 0.0003 g4 s30 °C to 175 °C0.02 % / 0.05 %200 mg
DIS 50-4HT50 g0.0001 g± 0.0003 g4 s30 °C to 250 °C0.02 % / 0.05 %200 mg
DIS 100-3120 g0.001 g± 0.003 g3 s30 °C to 175 °C0.03 % / 0.1 %2 g
DIS 100-3HT120 g0.001 g± 0.003 g3 s30 °C to 250 °C0.03 % / 0.1 %2 g

Technical details

These values apply to all models. Deviating values are given in the model table.

HeaterHalogen quartz glass heater, 400 W
Weighing systemForce compensation with overload protection
Weighing panØ 95 mm
Heating profilesStandard drying, rapid drying, gentle drying, step drying
Drying programmes (memory)40
Switch-off criteriaautomatic, time-controlled, manual
Memory1,000 measurement results, 1,000 sample names
DisplayGraphic LCD with backlight, digit height 15 mm
InterfacesRS-232 (standard), USB device
Adjustmentexternal
HousingABS plastic
Dimensions, fully assembled (W x D x H)211 x 342 x 187 mm
Net weightapprox. 5.0 kg
Power supplybuilt-in power supply, 230 V AC, 50 Hz, IEC power cable
Ambient temperature15 °C to 30 °C
Humidity25 % to 85 %
DAkkS calibration certificatecan be ordered separately, item no. 963-127
Warranty (manufacturer)36 months according to Kern, excluding wear parts

Scope of delivery

  • Kern DIS moisture analyser with halogen quartz glass heater
  • 50 sample pans
  • Working protective cover
  • IEC power cable

Calibration

A DAkkS calibration certificate (item no. 963-127) can be ordered separately. The confirmation of manufacturer specification (969-517) is only available in conjunction with the DAkkS calibration certificate. A factory calibration certificate (item no. 964-305) is available for temperature. The temperature calibration points are 100 °C and 175 °C; for the HT models, they are 25 °C, 100 °C and 250 °C.

Warranty

According to Kern & Sohn, these devices come with a 3-year warranty. Kern grants 3 years on products with a list price of €500 or more and 2 years on products below that and on components; wear parts such as batteries and rechargeable batteries are excluded. Warranty repairs are carried out by Kern or by a Kern dealer.

Downloads

Regulatory information

As electrical devices, the moisture analysers fall under the ElektroG. The manufacturer Kern & Sohn GmbH is registered under the WEEE registration number DE 80821200. How to return old appliances is described on the page Environment and disposal.

Individual requirements and procurement service

Aluminium sample pan (RH-A01), glass fibre filter circles (RH-A02, RH-A03), safety hoods (YBA-A23, YBA-A23S05), temperature calibration kit (DIS-A01), IEC power cable (822-903), USB 2.0 cable (DBS-A04), RS-232 interface cable (MLB-A05), thermal printer (YKC-01, YKH-01), dot matrix printers (YKT-01), RS232 Bluetooth adaptors (YKI-02), RS232 to Ethernet converters (YKI-10, YKI-11), RS232 Wi-Fi converter (YKI-12), OIML Class F1 test weights (326-06 at 50 g; for the 100-series models, also 326-07 at 100 g), a Safety Set (315-306-303 at 5 g and 50 g, Class F2) and the test for individual substances (965-243) are also available. Enquiries are handled via our procurement service.

Design: DIS 50-4 · 50 g
Power supply: Mains operation