WTW SenTix Sur pH surface electrode with flat membrane, glass shaft, S7 connector, 103646

ⓘ  This product description has been translated from German using AI. Despite careful review, we do not guarantee that the translation is error-free; in case of doubt, the German version shall prevail.

Product information "WTW SenTix Sur pH surface electrode with flat membrane, glass shaft, S7 connector, 103646"

A flat membrane does not measure within the sample, but on its surface. The WTW SenTix Sur is a pH surface electrode for moist samples such as leather, paper and skin. The membrane is flat and, according to WTW, impact-resistant. The shaft is made of glass, is 120 mm long and has a diameter of 12 mm. The measuring range is 2 to 13 pH; the sample temperature may range from 0 to 50 °C.

Flat membrane and annular gap

The flat membrane rests on the sample surface. An annular gap serves as the junction, an open junction that, according to WTW, is insensitive to contamination and easy to clean. The reference electrolyte is a polymer (Referid). The electrode is low-maintenance and is not refilled. The membrane resistance is less than 1 GΩ.

Typical applications according to WTW

Surface measurements on damp samples are the main application. WTW also cites applications in cosmetics, the pharmaceutical industry and biology.

  • Leather
  • Paper
  • Skin

When another SenTix model fits better

If the electrode is to penetrate the sample, for example in cheese, creams or toothpaste, WTW offers the SenTix Sp with a spear-shaped membrane and the same measuring range of 2 to 13 pH. For liquid samples in a beaker, there are glass electrodes with liquid electrolyte, such as the SenTix 60 with a range of 0 to 14 pH.

Connection and temperature

The SenTix Sur has an S7 plug head and is supplied without a cable. Depending on the measuring instrument, one of the WTW connection cables AS/DIN (1 m, 108110), AS/DIN-3 (3 m, 108112) or AS/BNC (1 m, 108114) fits. A temperature sensor is not fitted. The temperature range ends at 50 °C; this should be taken into account when measuring warm samples.

Storage and care

During measurement breaks, the electrode should be immersed in reference electrolyte (KCl 3 mol/l, Ag+-free). Overnight and for longer periods, it stays in the wetting cap, filled with the same electrolyte. According to WTW, storing it dry or in deionised water can cause permanent damage. If the cap has dried out, the electrode needs at least 24 hours in reference electrolyte. As with any pH electrode, strong acids and alkalis, hydrofluoric acid, organic solvents, oils, fats and proteins shorten its service life.

Technical details

TypepH surface electrode, analogue
Shaft materialGlass
Shaft length / Shaft diameter120 mm / 12 mm
ElectrolytePolymer (Referid), non-refillable
JunctionAnnular gap
Membrane shapeFlat
pH measuring range2 to 13
Temperature range0 to 50 °C
Membrane resistance< 1 GΩ
Temperature sensorNone
ConnectionS7 plug head
CableNo cable

Scope of delivery

  • 1 WTW SenTix Sur pH electrode with wetting cap
  • No connection cable in the scope of delivery

Accessories and procurement service

Connection cables (AS/DIN, AS/DIN-3, AS/BNC), KCl 3 mol/l reference electrolyte and buffer solutions for calibration are also available. The procurement service can answer any further questions regarding surface measurements.