Skip to main content Skip to search Skip to main navigation

Binder KB PRO 720 refrigerated incubator, extended temperature range –20 to 100 °C, 727 l, 9020-0436

Product information "Binder KB PRO 720 refrigerated incubator, extended temperature range –20 to 100 °C, 727 l, 9020-0436"

The Binder KB PRO 720 offers a usable volume of 727 litres with a temperature range of –20 °C to +100 °C. Whilst the KB series covers the standard incubation range of 0 °C to 70 °C, the KB PRO 720 is also suitable for applications requiring deep-freeze conditions or higher temperature profiles, offering a large volume for extensive sample series or multiple parallel test runs with different temperature profiles.

Compressor cooling with extended temperature range and R-290

The KB PRO 720 operates with an inverter-controlled refrigeration unit and the climate-neutral refrigerant R-290. At 37 °C, the spatial temperature deviation is ±0.2 K, the temporal deviation is ±0.1 K, and the maximum heat compensation is 1,000 W. Two internal and two external doors further organise the large interior space. The patented APT.line preheating chamber technology ensures homogeneous temperature distribution even when fully loaded. After the door has been open for 30 seconds, the set temperature of 37 °C is reached again in around 4 minutes; energy consumption is 315 Wh/h at 25 °C and 330 Wh/h at 37 °C.

Typical applications

The extended temperature range makes the KB PRO 720 a versatile tool wherever conventional refrigerated incubation reaches its limits; with this unit size, this is particularly the case with high sample volumes. Typical applications:

  • Microbiological incubation and cold stress tests with large batches of bacterial, yeast and fungal cultures
  • Stability testing in pharmaceutical quality control across a wide temperature range
  • Sensory and shelf-life tests in food analysis with high sample throughput, even under deep-freeze conditions
  • Materials testing and research with defined temperature profiles from minus 20 °C to 100 °C
  • Test laboratories with multiple parallel test runs and varying temperature requirements

Features and durability for everyday laboratory use

The interior of the KB PRO 720 is made entirely of stainless steel; the inner doors are made of metal oxide-coated safety glass and allow visual inspection of the samples without opening the outer doors. A door heater prevents condensation from forming. Two stainless steel shelf racks with telescopic rails are included as standard; the interior can be expanded to accommodate up to 16 shelf racks, each with a load capacity of 40 kg, with a maximum total load of 200 kg. A feed-through with an 80 x 45 mm opening and silicone plug allows external sensor or measurement cables to be fed in. To ensure process reliability, an independent Class 3.3 temperature selector monitor in accordance with DIN 12880, featuring visual and audible alarms, is integrated; an internal data logger records the measured values and can be read out via USB and Ethernet. The APT-COM Basic Edition management software is included as standard, whilst a device self-test enables a comprehensive status analysis.

Technical details

Temperature range–20 °C to +100 °C
Useful volume727 l
Internal dimensions (W x H x D)1,011 x 1,250 x 575 mm
Temperature variation within the chamber (at 37 °C)± 0.2 K
Temperature variation over time (at 37 °C)± 0.1 K
Max. thermal compensation (at 37 °C)1,000 W
Recovery time after door opening (30 s, at 37 °C)4 min
Energy consumption at 25 °C / 37 °C315 Wh/h / 330 Wh/h
RefrigerantR-290 (climate-neutral)
Connected load1.7 kW
Rated voltage208 to 240 V
External dimensions (W x D x H)1,285 x 900 x 1,951 mm
Unit weight (net, empty)296 kg
Doors (internal/external)2 / 2
Shelf grids (standard/max.)2 / 16, stainless steel with telescopic rails
Max. load capacity per drawer40 kg
Max. total load200 kg
Feed-through80 x 45 mm, with silicone plugs
Certification: Temperature selector switchClass 3.3 in accordance with DIN 12880, with visual and audible alarm

Contents

  • 1x Binder KB PRO 720 refrigerated incubator (9020-0436)
  • 2x stainless steel shelves with telescopic rails
  • BINDER Multi Management Software APT-COM Basic Edition
  • Operating instructions

Plant light, ICH and CO2 modules for the KB PRO 720

Binder KBF PRO 720 with ICH light moduleBinder KBF PRO 720 with LED plant light module
Pictured: KBF PRO 720 with ICH light module (left) and with LED plant light module (right). The modules are accessories.

The KB PRO 720 accepts three modules, the LED plant light module (8012-2439), the ICH light module (8012-2442, size 720) and the CO2 gassing module (8012-2140 for 0 to 1 vol.%, 8012-2138 for 0 to 20 vol.%). In LED operation, the PAR photon flux density is 1,080 µmol/m²/s at 25 cm, 740 µmol/m²/s at 50 cm and 560 µmol/m²/s at a distance of 75 cm, and the temperature range narrows to −10 °C to +60 °C. This series does not offer humidity control at all. With the ICH module, the chamber reaches 42 klx VIS illuminance and 51 W/m² UVA irradiance, a complete ICH Q1B test is finished in 33 hours, usable between 5 °C and 60 °C. The CO2 module does not change the temperature range, which remains at −20 °C to +100 °C. The shelf set with U-rails for the LED strips carries item number 8012-2391.

The modules are accessories and not included with the chamber. You will find them with prices under "Matching modules" on this page; more on our Binder page.

Customised requirements & procurement service

Genuine accessories are available for the KB PRO 720, such as additional shelf racks, a flexible Pt100 internal sensor for displaying or controlling the sample temperature, further silicone feed-throughs in various sizes, or the APT-COM management software (Professional Edition) for connecting to a data network. Calibration certificates and maintenance contracts can also be arranged if required. Do you require additional accessories, a bespoke configuration or assistance in selecting the right cooling incubator for your application? LT Laborhandel will be happy to assist you with procurement and configuration via our procurement service.

Volume: 727 l
Skip product gallery

Matching modules

Promo
Binder LED-Pflanzenlichtmodul 8012-2439, Lichtsteuerbox mit Programm-Controller
Binder LED plant light module for climate chambers and refrigerated incubators, 16 LED light bars, 8012-2439
Herstellernummer: 8012-2439

A Binder constant-climate cabinet or refrigerated incubator can be converted into a plant growth cabinet using the 8012-2439 LED plant light module, without the need to purchase a separate unit. The module produces lighting conditions that closely resemble natural sunlight and fits into cabinets from the KB, KB PRO, KBF and KBF PRO series with development stage E7. Installation is carried out on site by the customer or a service partner. Design and lighting control The Binder LED plant lighting module consists of 16 LED light strips, which can be positioned individually within the cabinet, and a lighting control box with its own programme controller. Each light strip has three separately controllable colour channels: daylight-like cool white (SunLike, 6,500 K), daylight-like warm white (SunLike, 2,700 K) and dark red phytochrome-controlled light (730 nm), which specifically activates the plant’s phytochrome system. All three channels can be dimmed independently of one another from 0% to 100%. Real-time programming is available for daily routines, either as a cyclical daily programme or individually for each day of the week, supplemented by a twilight function that gently fades the light on and off over a one-hour ramp rather than switching it on or off abruptly. An integrated ventilation function feeds 0 to 500 litres of ambient air per hour into the chamber, thereby maintaining the CO₂ level close to the ambient value required by plants for photosynthesis. The light module is powered by 120 V to 240 V at 50 or 60 Hz, regardless of the mains voltage of the climate cabinet itself. Light output by cabinet size The light intensity depends on the cabinet size and the distance from the light source. With 100 per cent white light and all 16 light strips evenly distributed on a single plane, size 260 achieves a photosynthetically active radiation (PAR) of 1,660 µmol/m²/s at a distance of 25 cm. Size 470 achieves 1,540 µmol/m²/s at a distance of 25 cm, 990 µmol/m²/s at 50 cm and 730 µmol/m²/s at 75 cm. For the 720 size, the figures are 1,080, 740 and 560 µmol/m²/s at the three distances; for the 1060 size, they are 880, 610 and 480 µmol/m²/s respectively. In the largest available size, 1600 – which is only available for the KB PRO and KBF PRO models – the values are 640, 470 and 380 µmol/m²/s. Distributing the 16 light strips across several levels reduces the light output accordingly: to around 50 per cent of the initial value with two levels, and to around 25 per cent with four levels. The maximum temperature during lighting operation is 60 °C for all sizes. Order mounting grids separately The LED plant light module does not include any grids for mounting the light strips. However, LT Laborhandel supplies insert grid sets with telescopic rails (size 260/470: 8012-2384, size 720: 8012-2385, size 1060: 8012-2386, size 1600: 8012-2387) as well as insert grille sets with U-rails, which Binder expressly recommends for mounting the light strips (size 260/470: 8012-2390, size 720: 8012-2391, size 1060: 8012-2392, size 1600: 8012-2393). The number of additional grilles required depends on the desired lighting configuration. For the basic configuration without an additional level, the two standard grids supplied with the cabinet are sufficient; for one or two light levels, at least one additional set is recommended; for full distribution across several levels, up to six sets are required. An additional grid between the light bars and the sample material also creates a gap of around 10 cm, thereby reducing thermal effects on the samples. Typical applications Plant growth experiments with day-night simulation via real-time programming Germination and seedling experiments with reduced light intensity on a single light bar level Photoperiodism and phytochrome studies via the separately dimmable dark red channel Retrofitting existing Binder E7-generation climate chambers and refrigerated incubators to create plant growth chambers, without the need for additional equipment If a higher CO₂ concentration is also required, for example to accelerate plant growth, the LED plant light module can be combined with a CO₂ injection module. According to the operating instructions, both accessories can operate simultaneously within the same cabinet, unlike the ICH light module, which occupies the CO₂ connection. Technical details Configuration16 LED light strips, individually positionable Colour channels per light stripCool white 6,500 K, warm white 2,700 K, dark red 730 nm Dimming range per channel0% to 100% Ventilation function0 to 500 litres of ambient air per hour PFR PAR, size 260 (25 cm)1,660 µmol/m²/s PFR PAR, size 470 (25/50/75 cm)1,540 / 990 / 730 µmol/m²/s PFR PAR, size 720 (25/50/75 cm)1,080 / 740 / 560 µmol/m²/s PFR PAR, size 1060 (25/50/75 cm)880 / 610 / 480 µmol/m²/s PFR PAR, size 1600 (25/50/75 cm)640 / 470 / 380 µmol/m²/s Maximum temperature during operation60 °C Compatible series (development stage E7)KB, KB PRO, KBF, KBF PRO Power supply for light module120 V to 240 V, 50/60 Hz Contents of delivery 16 LED light strips with retaining clips for clipping into standard slotted grilles Lighting control box with programme controller Mains cable Power distribution cable for the LED light strips PU tubing for the air supply Operating instructions Insert grilles for mounting the light strips are not included. These must be ordered as separate accessories; see the section above. Customised requirements & procurement service LT Laborhandel can assist in putting together the right combination of climate chamber, lighting module and insert grilles, for example if the existing size or series is unclear or if multiple lighting levels are planned. The procurement service is also available for other Binder accessories such as CO₂ gas injection modules, water supply systems or sample temperature control sensors. Details can be found on our Services page.

Manufacturer's RRP: €5,278.00
Promo
Binder ICH light module for climate chambers size 720, photostability test to ICH Q1B, 8012-2442
Binder ICH light module for climate chambers size 720, photostability test to ICH Q1B, 8012-2442
Herstellernummer: 8012-2442

Anyone who needs to carry out photostability tests in accordance with ICH Q1B using a larger sample size will find that the 720 size offers more usable volume than the smaller climate chambers. The Binder ICH light module 8012-2442 is designed specifically for this size and fits into climatic chambers from the KB PRO, KBF and KBF PRO series with development stage E7. Light cassette technology The module consists of a power supply unit for the side panel of the unit, as well as two light cassettes: one for visible light (VIS) and one for UVA radiation. The VIS cassette utilises LED technology and a spectrum similar to sunlight (SunLike Cool White, 6,500 K), whilst the UVA cassette uses a special fluorescent tube (Narva PUVA), as, according to the manufacturer, current LED technology is not yet capable of reproducing this spectrum in an ICH-compliant manner. Together, both cassettes generate the CIE D65 spectrum required by ICH Q1B, which corresponds to daylight grey under overcast skies. Control is via a data cable connected to the climate chamber; the irradiation duration is set as a time-interval programme on the programme controller. The light module is powered by 120 V to 240 V at 50 or 60 Hz. Lighting values and test duration for size 720 In size 720, at full light output, the module achieves a VIS illuminance of 42 klx and a UVA irradiance of 51 W/m², both measured using spherical evaluation at a distance of 12 cm from the light source. These higher values compared with the smaller sizes 260 and 470 reduce the test duration: A complete ICH-Q1B test with the required light doses of 1.2 Mlx·h of visible light and 200 Wh/m² of UVA can be completed in 29 hours for the VIS component and 4 hours for the UVA component, totalling 33 hours. For flat samples, such as printed packaging materials, the test duration is extended by a factor of 1.5 according to the manufacturer, as the spherical evaluation was originally calibrated for three-dimensional test specimens. For pharmaceutical companies with a high sample throughput, this represents a significant time saving compared with the smaller models, without altering the required light doses in any way. Control system without automatic light dose control This version has no light sensors and operates with a conventional timer: the required irradiation time is calculated once on the basis of the known light values and stored as a programme. For automatic switch-off once the target dose has been reached, including continuous measurement, the LQC variant 8012-2444 is available for the same model. In both cases, installation is carried out on site by the customer or a service partner, without the need to return the climate chamber. Typical applications Photostability testing of larger batches or sample series in accordance with ICH Q1B Stability studies for marketing authorisation dossiers of new active substances Light stress tests on primary packaging and packaging materials Combined climate and light testing in a single, larger cabinet According to the operating instructions, a CO₂ gas injection module cannot be operated simultaneously with the ICH light module, as both occupy the same interface on the climatic chamber. An LED plant light module of the same size, however, can be operated without conflict on a second climatic chamber of the same series, provided that a further unit is available. Technical details Compatible size720 Compatible series (development stage E7)KB PRO, KBF, KBF PRO Temperature range for KB PRO 720 with light5 °C to 60 °C, unregulated humidity Temperature range for KBF 720 with light20 °C to 50 °C, 20% to 50% relative humidity Temperature range KBF PRO 720 with light10 °C to 50 °C, 10% to 80% relative humidity VIS illuminance42 klx UVA irradiance51 W/m² ICH Q1B test duration29 + 4 = 33 hours Light spectrum VIS cassetteLED SunLike Cool White, 6,500 K Light spectrum, UVA cassetteNarva PUVA fluorescent tube Power supply for light module120 V to 240 V, 50/60 Hz Contents VIS light cassette with connection cable UVA light cassette with connection cable U-rails for mounting the light cassettes in the climate chamber Power supply box with ballasts/power supplies Profile rails and bolts for securing the power supply box to the climate chamber Data cable Mains cable Operating instructions The light sensors for light dose control are not included in the scope of delivery; these are provided as standard with the ICH-LQC light module 8012-2444. Customised requirements & procurement service LT Laborhandel’s procurement service can advise you on choosing between the ICH light module and the ICH-LQC light module for size 720, as well as on other accessories for constant-climate cabinets. Details can be found on our services page.

Manufacturer's RRP: €4,430.00
Promo
Binder ICH-LQC light module for climate chambers size 720, with light dose control to ICH Q1B, 8012-2444
Binder ICH-LQC light module for climate chambers size 720, with light dose control to ICH Q1B, 8012-2444
Herstellernummer: 8012-2444

In size 720, photostability tests in accordance with ICH Q1B can be carried out with a larger sample volume, and, with the Binder ICH-LQC light module 8012-2444, can also be completed using a measured light dose rather than a calculated one. The module is designed for climate chambers of the KB PRO, KBF and KBF PRO series with development stage E7 in this size. Light dose control via two spherical sensors In addition to the VIS and UVA light cassettes, which together generate the CIE D65 spectrum in accordance with ICH Q1B, the ICH-LQC light module contains two additional spherical light sensors. The VIS sensor measures the illuminance of visible light, whilst the UVA sensor measures the irradiance of the UVA component; both have a spherical response characteristic for three-dimensional test specimens. The light dose to be achieved is stored in the climate chamber’s programme controller – typically 1.2 Mlx·h of visible light and 200 Wh/m² of UVA – and both light cassettes switch off independently of one another as soon as their respective doses have been reached. For flat samples, such as blister packs or printed packaging materials, the light dose must be increased by a factor of 1.5 according to the manufacturer, as the spherical sensors were originally calibrated for three-dimensional test specimens. Higher light output thanks to the larger size In size 720, the module achieves a VIS illuminance of 42 klx and a UVA irradiance of 51 W/m² at full light output, both measured using spherical evaluation at a distance of 12 cm from the light source. These higher values compared with the 260 and 470 sizes reduce the calculated test duration to 29 hours for the VIS component and 4 hours for the UVA component, totalling 33 hours. The LQC sensor system continuously monitors the actual dose achieved, rather than relying solely on this calculated duration. This is particularly relevant for larger sample quantities, where individual areas of the chamber may be illuminated slightly differently. When the LQC version is worthwhile Anyone requiring a documented, measured light dose rather than pure time-based programming – for example, for regulatory testing with a requirement for evidence – is better served by the LQC version than by the simpler ICH light module 8012-2442 for the same size. Both variants are installed on-site by the customer or a service partner and controlled via the climate chamber’s existing programme controller; no additional external control unit is required. Particularly in the case of the 720 size, with its larger usable volume, the LQC sensor system offers the added benefit that several sample batches can be tested simultaneously under a verifiably uniform light dose, rather than having to recalculate the irradiation time for each batch. For a pharmaceutical company with several parallel regulatory approval projects, this significantly reduces the coordination effort required between test series. Typical applications ICH Q1B photostability tests with a documented, measured light dose Regulatory stability testing with a larger sample size Photostability tests on flat samples with an adjusted dose specification Recurring test series with high demands on reproducibility According to the operating instructions, a CO₂ gas injection module cannot be operated simultaneously with the ICH-LQC light module, as both occupy the same interface on the climate chamber. Technical details Compatible model720 Compatible series (development stage E7)KB PRO, KBF, KBF PRO Temperature range for KB PRO 720 with light5 °C to 60 °C, unregulated humidity Temperature range for KBF 720 with light20 °C to 50 °C, 20% to 50% relative humidity Temperature range KBF PRO 720 with light10 °C to 50 °C, 10% to 80% relative humidity VIS illuminance42 klx UVA irradiance51 W/m² ICH Q1B test duration (calculated)29 + 4 = 33 hours Light sensorsSpherical VIS sensor, spherical UVA sensor Dose factor for flat specimens1.5 times the light dose Light spectrum of VIS cassetteSunLike cold white LED, 6,500 K Light spectrum of the UVA cassetteNarva PUVA fluorescent tube Power supply for light module120 V to 240 V, 50/60 Hz Contents VIS light cassette with connection cable UVA light cassette with connection cable U-rails for mounting the light cassettes in the climate chamber Power supply box with ballasts/power supplies Profile rails and bolts for securing the power supply box to the climate chamber Data cable Mains cable Operating instructions Spherical VIS light sensor for light dose control Spherical UVA light sensor for light dose control Customised requirements & procurement service LT Laborhandel’s procurement service can advise you on choosing between the ICH light module and the ICH-LQC light module for the 720 size, as well as on other accessories for constant-climate cabinets. Details can be found on our services page.

Manufacturer's RRP: €8,456.00
Promo
Binder CO2-Begasungsmodul
Binder CO2 gassing module for climate chambers, control range 0 to 1 vol.%, NDIR sensor, 8012-2140
Herstellernummer: 8012-2140

In a closed climate chamber, plants consume CO₂ during photosynthesis, which can cause the concentration to fall below the natural ambient level of around 0.04 vol.%. The Binder CO₂ gas supply module 8012-2140, on the other hand, maintains the concentration precisely between 0 and 1 vol.% and is therefore designed for plant growth applications where precise, low CO₂ control is required. It is an accessory for Binder KB PRO series refrigerated incubators and KBF and KBF PRO series constant-climate chambers, all of which are at development stage E7. Measurement and control CO₂ measurement is carried out by an infrared measuring cell with an NDIR sensor (non-dispersive infrared sensor) and continuous control to a reference value. According to the manufacturer, this measurement method offers fast response times and high selectivity for CO₂; interference and ageing effects in the measurement system are largely ruled out, meaning that the system remains virtually drift-free between calibrations. The control process itself is managed entirely via the MB2 controller of the climate chamber; the gas injection module cannot be operated without being connected to the climate chamber. To ensure the CO₂ concentration is displayed correctly in vol.%, the altitude of the installation site must also be entered into the controller. As this control range uses a separate CO₂ sensor to that of the 0–20 vol.% variant, the controller function ‘CO₂ measurement range 0…1 vol.%’ must also be activated to ensure the measured value is converted correctly. Connections and interfaces The CO₂ gas injection unit has a 4 to 20 mA analogue output for the CO₂ concentration, via which the measured value can be transmitted to external data acquisition systems or recording devices. The gas connection is made via a quick-release coupling on the gas cylinder, whilst the data connection to the climate chamber is via a patch cable. A CO₂ gas cylinder is not included in the scope of delivery and must be procured separately. Combination with other accessory modules According to the operating instructions, the CO₂ gas module can be operated simultaneously with the LED plant light module; both accessories complement each other for comprehensive plant growth applications with controlled lighting and CO₂ concentration within the same cabinet. However, simultaneous operation with the ICH light module or the ICH-LQC light module is not possible, as both accessories occupy the same interface on the climate chamber. Anyone requiring both configurations – for example, CO₂ gasation and the ICH photostability test – should plan for two separate climate chambers. For pure plant growth experiments with LED lighting and CO₂ enrichment, however, a single, suitably equipped climate chamber is sufficient. Operation and maintenance A fine gas filter in the gas injection box can be replaced as required to ensure that measurement accuracy is maintained throughout the operational life of the unit. Two CO₂ warning stickers are included in the scope of delivery and must be affixed to a clearly visible position on the climate chamber, as special safety instructions must be observed when handling CO₂ in enclosed spaces. Before initial commissioning, the correct CO₂ outlet pressure of the gas supply and the tightness of the connections must be checked. A connection panel on the rear of the incubator or climate chamber houses the data connection to the gas supply unit, ensuring that the cabling remains neatly organised. Typical applications CO₂ enrichment for accelerated plant growth in combination with the LED plant light module Photosynthesis and growth studies with a defined, low CO₂ concentration Offsetting CO₂ consumption in densely planted, enclosed climate chambers Botanical and agricultural research series with reproducible climatic conditions The CO₂ injection box is installed on the prepared side panel of the climate chamber and connected to the main unit via a data cable, gas hose and sensor cable. Technical details CO₂ control range0 vol.% to 1 vol.% Minimum concentrationAmbient value (approx. 0.04 vol.%) Measuring principleNDIR sensor (non-dispersive infrared sensor) Analogue output4 to 20 mA DC Compatible series (development stage E7)KB PRO, KBF, KBF PRO Compatible sizes260, 470, 720, 1060, 1600 Controlvia MB2 unit controller; operation not possible without a climate cabinet Power supply for fumigation box24 V, via the supplied mains adaptor Scope of supply CO₂ fumigation box CO₂ sensor with cable 2 CO₂ warning stickers 24 V mains adaptor with adaptors Operating instructions 2 mounting rails with screws (4 × M8, 2 × M5) 2 silicone foam plugs for the slotted hole feed-through Patch cable for the data connection between the gas supply box and the device Hose set for the gas connection The CO₂ gas cylinder and pressure regulator are not included; these must be purchased separately. Customised requirements & procurement service LT Laborhandel provides procurement support when choosing between the CO₂ control ranges of 0 to 1 vol.% and 0 to 20 vol.%, as well as when combining the system with an LED plant lighting module or water supply systems. Details can be found on our services page.

Manufacturer's RRP: €3,925.00
Promo
Binder CO2-Begasungsmodul
Binder CO2 gassing module for climate chambers, control range 0 to 20 vol.%, NDIR sensor, 8012-2138
Herstellernummer: 8012-2138

Some applications require significantly higher CO₂ concentrations than those found in natural plant growth, such as cell culture work, which traditionally requires a CO₂ incubator. With its control range of 0 to 20 vol.%, the Binder CO₂ gas supply module 8012-2138 covers precisely this broader spectrum and additionally equips a Binder KB PRO series refrigerated incubator or a KBF and KBF PRO series constant-climate chamber (development stage E7) with CO₂ control. Measuring principle and operation Like the version with the narrower control range, this module also measures the CO₂ concentration using an infrared measuring cell with an NDIR sensor and provides continuous control to a reference value. The method has fast response times and, according to the manufacturer, remains virtually drift-free between calibrations, as interference and ageing effects in the measuring system are largely ruled out. The CO₂ control is managed entirely via the MB2 controller of the climate chamber; standalone operation without a connected climate chamber is not intended. To ensure the concentration is displayed correctly in vol.%, the sea-level altitude of the installation site must be entered into the controller so that differences in air pressure do not distort the measurement. Connections and gas supply The CO₂ gas supply box has an analogue output of 4 to 20 mA for the CO₂ concentration, via which the measured value can be transmitted to an external data acquisition system or a recording device. The gas connection is made via a quick-release coupling, connected to a CO₂ gas cylinder, which is not included in the scope of delivery and must be procured separately. The data connection between the gas injection box and the climate chamber is established via a patch cable supplied with the unit; a fine gas filter in the gas injection box can be replaced if necessary. Cannot be combined with the light modules According to the operating instructions, the CO₂ gas injection module, with a control range of 0 to 20 vol.%, cannot be operated simultaneously with the ICH light module or the ICH-LQC light module, as both occupy the same interface on the climate chamber. Simultaneous operation with the LED plant light module, however, is possible, even though the wide control range up to 20 vol.% is rarely utilised in practice for pure plant growth applications and is intended more for other gas injection tasks. Those who require only low, plant-specific concentrations of up to 1 vol.% are generally better served by the more precise control range of the 8012-2140 variant. Operation and maintenance The fine gas filter in the gas supply box can be replaced as required to ensure that measurement accuracy is maintained throughout the operational life of the unit. Two CO₂ warning stickers are included in the scope of delivery and should be affixed to a clearly visible position on the climate chamber, as special safety precautions must be observed when handling CO₂ in enclosed spaces. Before initial commissioning, the correct CO₂ outlet pressure of the gas supply and the tightness of all connections must be checked, particularly given the significantly wider control range of up to 20 vol.% in this case. A connection panel on the rear of the incubator or climate chamber houses the data connection to the gas supply unit, ensuring that the cabling remains neatly organised. Typical applications Upgrading a climate chamber to function as a CO₂ incubator, for example for cell culture work Gassing experiments with higher CO₂ concentrations beyond purely botanical applications Processes that require a documented, constant CO₂ atmosphere to be recorded via an analogue output Research applications with varying CO₂ setpoints across the entire control range Installation takes place on the prepared side panel of the climate chamber, connected via a data cable, gas hose and sensor lead. Technical details CO₂ control range0 vol.% to 20 vol.% (factory setting) Minimum concentrationAmbient value (approx. 0.04 vol.%) Measurement principleNDIR sensor (non-dispersive infrared sensor) Analogue output4 to 20 mA DC Compatible series (development stage E7)KB PRO, KBF, KBF PRO Compatible sizes260, 470, 720, 1060, 1600 Controlvia MB2 unit controller; operation not possible without a climate cabinet Power supply for fumigation box24 V, via the supplied mains adaptor Scope of supply CO₂ fumigation box CO₂ sensor with cable 2 CO₂ warning stickers 24 V mains adaptor with adaptors Operating instructions 2 mounting rails with screws (4 × M8, 2 × M5) 2 silicone foam plugs for the slotted hole feed-through Patch cable for the data connection between the gas supply box and the device Hose set for the gas connection The CO₂ gas cylinder and pressure regulator are not included; these must be purchased separately. Customised requirements & procurement service LT Laborhandel’s procurement service can assist you in choosing between the CO₂ control ranges of 0 to 20 vol.% and 0 to 1 vol.%, as well as with other accessories for constant-climate cabinets and refrigerated incubators. Details can be found on our services page.

Manufacturer's RRP: €3,179.00