Product information "Brand Dispensette S Organic bottle-top dispenser, Analog, 1-10 ml, without Recirculation Valve, DE-M, 4630140"

For acid digestion using a fixed volume of concentrated hydrochloric or nitric acid, the setting does not need to be checked again for each sample. Once set, the Brand Dispensette S Organic Analog (1 to 10 ml) remains at the selected value. For volumes of 1-10 ml, the graduation is 0.2 ml, the accuracy is 0.5 %/50 µl, and the coefficient of variation is 0.1 %/10 µl.

Typical applications

The colour code ‘Yellow’ identifies the Organic series within the Brand range. For the 1-10 ml range, typical applications in practice include:

  • fixed dosing of concentrated hydrochloric acid in standardised digestion procedures
  • consistent addition of dichloromethane during liquid-liquid extraction
  • recurring addition of THF during standardised synthesis steps

Analog version and selection guide

The volume is set using the volume adjustment screw with indicator arrow: loosen the screw by three-quarters of a turn, move the indicator arrow to the scale value, then tighten the screw again. Once set, the value remains fixed until deliberately adjusted. Analog instruments can be adjusted tool-free via Easy Calibration. Bottle adaptors for GL 24-25, GL28/S 28, GL 32-33, GL 38 and S 40 are included; the filling tube can be adjusted to 125-240 mm.

For a digestion protocol requiring a consistent acid volume, the mechanical Analog version is sufficient. If the volume varies between sample series, the Digital version with a counter is more practical.

Materials and autoclavability

Parts in contact with the medium are made of borosilicate glass, Al2O3 ceramic, ETFE, FEP, PFA, PTFE, tantalum and PP (screw cap/bottle adapter). The device is autoclavable under the following conditions: 121 °C, 2 bar, for at least 15 minutes according to DIN EN 285, with a cooling time of approx. 2 hours in a closed autoclave. The use of tantalum as the valve spring material, together with the fluoroplastics ETFE, FEP, PFA and PTFE, makes the Organic series suitable for use with organic solvents and certain concentrated acids; however, it is not suitable for alkalis or saline solutions.

The Dispensette S Organic is not suitable for alkalis, salt solutions, carbon disulphide, explosive liquids or media that attack borosilicate glass, such as hydrofluoric acid. For alkalis and salt solutions, the Dispensette S in red (standard) is the suitable series; for trace analysis with a selectable valve spring made of platinum-iridium or tantalum, the Dispensette S Trace Analysis is the suitable series.

Installation and switching to a different stock bottle are carried out using the supplied mounting key and the telescopic filling tube, which can be adjusted to the height of the respective bottle. The bottle adapter included in the scope of delivery fits the relevant GL45-compatible thread, allowing the dispenser to be used regardless of the specific stock bottle.

Accuracy at a glance

For 1-10 ml: graduation 0.2 ml, accuracy 0.5 %/50 µl, coefficient of variation 0.1 %/10 µl. The DE-M marking in the product name refers to conformity assessment under German weights and measures law; it does not replace a calibration certificate.

Other volumes in the Analog version

The Analog version of the Dispensette S Organic covers further ranges in addition to 1-10 ml:

VolumeAccuracyMPN (without RV)
0.5-5 ml0.5 %/25 µl4630130
2.5-25 ml0.5 %/125 µl4630150
5-50 ml0.5 %/250 µl4630160
10-100 ml0.5 %/500 µl4630170

Frequently asked questions

Can this version be retrofitted without a recirculation valve?
According to the manufacturer, the version with a recirculation valve has its own order number and comes with its own discharge tube including a recirculation tube; it is not a retrofit option for the version without a recirculation valve.

How is the Analog version adjusted in the 1-10 ml range?
Easy Calibration allows this to be done without tools; any adjustment deviating from the factory setting is indicated by a red mark.

Technical details

VersionAnalog
Recirculation valveNo
Volume range1-10 ml
Graduation0.2 ml
Accuracy (R)0.5 %/50 µl
Coefficient of variation (CV)0.1 %/10 µl
Colour codeyellow (Organic)
Valve springTantalum
Materials in contact with the mediumBorosilicate glass, Al2O3 ceramic, ETFE, FEP, PFA, PTFE, tantalum, PP (screw cap/bottle adapter)
AutoclavableYes, 121 °C, 2 bar, at least 15 minutes according to DIN EN 285, cooling time approx. 2 hours in a closed autoclave
ThreadGL45
Adapter/filling tubeAdapters: GL 24-25, GL28/S 28, GL 32-33, GL 38, S 40; filling tube 125-240 mm

Scope of delivery

  • 1 Dispensette S Organic bottle-top dispenser for GL45-threaded bottles
  • Discharge tube without recirculation valve
  • Telescopic filling tube
  • Mounting key
  • Bottle adaptor suitable for the relevant volume range
  • Quality certificate
  • Operating instructions

Differentiation within the Dispensette S range

For the medium being dispensed here, the yellow Organic series is suitable. For alkalis or most other acids, the red Standard Dispensette S would be the suitable choice instead. For trace analysis, with a choice of valve spring made from platinum-iridium or tantalum, the Dispensette S Trace Analysis is available separately.

Individual requirements and procurement service

Any additional requirements for bottle adaptors, replacement valves or further volume ranges of the Dispensette S Organic can be arranged via our procurement service.

Display type: Analogue
Operation: mechanical
Recirculation valve: without recirculation valve
Volume range: 1 ml - 10 ml
Volume setting: variable