Skip to main content Skip to search Skip to main navigation

Memmert SN110plus Hot-Air Steriliser, 108 L, Natural Convection, TwinDISPLAY – Class IIb medical device

Product information "Memmert SN110plus Hot-Air Steriliser, 108 L, Natural Convection, TwinDISPLAY – Class IIb medical device"

The Memmert SN110plus hot-air steriliser sterilises medical materials using dry heat at atmospheric pressure – as a Class IIb medical device (in accordance with MDD 93/42/EEC), it is suitable for hospital laboratories, microbiological and pharmaceutical laboratories, as well as larger medical practices with regular sterilisation requirements. With an internal capacity of 108 litres and a temperature range of +20 °C to +250 °C, the unit is particularly suitable for the reliable sterilisation of multiple batches of instruments, vessels or powders during day-to-day practice or laboratory operations. The combination of certified medical device classification and a validatable sterilisation process supports the requirements for traceability and documentation that are standard in the quality management systems of medical practices, hospitals and laboratories.

Natural convection instead of forced air circulation

Unlike the Memmert SF series, the SN110plus operates using natural ventilation rather than an electronically controlled recirculation fan. Preheated fresh air is introduced via an electronically adjustable exhaust air damper, whilst an exhaust air outlet with an adjustable throttle valve ensures controlled air exchange – entirely without any active air movement inside the chamber. This makes the SN series the preferred choice wherever powders, granules or light, exposed instruments could be stirred up or contaminated by a forced air flow.

Precise temperature control with SetpointWAIT

The TwinDISPLAY version SN110plus provides additional safeguards for this process: two independent Pt100 sensors using a 4-wire measurement method monitor the temperature in tandem – if one sensor fails, the second seamlessly takes over control. The SetpointWAIT function only starts the sterilisation time once the setpoint temperature has actually been reached – ensuring validatable, reproducible cycles, as required for quality management documentation in dental and clinical hygiene, as well as in microbiological and pharmaceutical laboratories.

Safety and Features

  • Electronic temperature monitoring plus mechanical temperature limiter, protection class 1 (DIN 12880)
  • Self-diagnostic system for fault detection, visual and audible alarms
  • AutoSAFETY: additional integrated over- and under-temperature monitoring
  • Controller: TwinDISPLAY, adaptive multifunctional digital PID microprocessor controller with high-resolution TFT colour graphic display
  • Digital countdown timer with target time setting, adjustable from 1 minute to 99 days

Operation, interfaces and documentation

Programmes can be managed and log data from the data logger read out and archived via the Ethernet interface or, optionally, via a USB stick using the AtmoCONTROL software – the programme sequence remains stored even after a power cut. This significantly facilitates comprehensive documentation in quality management, a criterion that is particularly important for validatable sterilisation processes in hospitals and laboratories. The built-in self-diagnostic system reports faults at an early stage via visual and audible alarms, so that deviations from the target process are not only noticed at the end of a sterilisation cycle, but can be detected and documented immediately.

Typical areas of application

  • Sterilisation of instruments in medical practices and hospital hygiene
  • Sterilisation of powders, granules and bulk materials without air movement inside the chamber
  • Reprocessing of laboratory glassware and open vessels in microbiological and pharmaceutical laboratories
  • Validatable, documented sterilisation cycles in quality management

This allows a large batch of instruments or vessels to be sterilised reliably and with full traceability in a single operation whilst clinical or laboratory work is ongoing, without sensitive items being displaced by air flow.

Technical details

Internal chamber volume108 litres
Internal dimensions (W×H×D)560 x 480 x 400 mm
External dimensions (W×H×D)745 x 864 x 584 mm (+56 mm depth for door handle)
Temperature range+20 °C to +250 °C
Temperature accuracy±0.1 °C (up to 99.9 °C) / ±0.5 °C (from 100 °C)
Sensors2 Pt100 sensors, DIN Class A, in a 4-wire configuration (cross-monitoring)
ConvectionNatural ventilation (no circulation fan)
Fresh air/exhaust airElectronically adjustable exhaust air damper, exhaust air outlet with adjustable throttle valve
ControlTwinDISPLAY, digital PID microprocessor controller, TFT colour graphic display
TimerDigital, 1 minute to 99 days
Safety functionSetpointWAIT (start only once the setpoint temperature has been reached)
Temperature monitoringElectronic + mechanical temperature limiter (DIN 12880 Class 1)
Max. load175 kg total / 20 kg per shelf
Max. number of shelves5
InterfacesEthernet, USB (with AtmoCONTROL software)
Voltage/Power230 V/50/60 Hz at approx. 2800 W or 115 V/50/60 Hz at approx. 1800 W
Installation requirementsMin. 15 cm clearance from the rear wall, 20 cm clearance from the ceiling, 5 cm clearance at the sides; max. installation height 2,000 m
Environmental conditions+5 °C to +40 °C, max. 80% relative humidity (non-condensing)
Packaging dimensions830 x 1050 x 800 mm
Net weight/Gross weightapprox. 74 kg / approx. 99 kg
CertificationsCE (LGA, 0197), VDE, UL
Medical device classificationClass IIb in accordance with MDD 93/42/EEC

Contents of delivery

  • 1x Memmert SN110plus hot-air steriliser (basic unit)
  • 2x stainless steel wire racks (electropolished)
  • 1x factory calibration certificate (at +160 °C)

Regulatory information

WEEE registration number (manufacturer Memmert): DE 66812464. As an electrical appliance, the SN110plus falls under the ElektroG (WEEE Directive) – registration is held by the manufacturer Memmert, as they are the first to place the appliance on the German market, not by LT Laborhandel as the distribution partner.

Customised requirements & procurement service

Do you require additional accessories such as further stainless steel grids, a bespoke calibration or spare parts? Our procurement service will be happy to assist you with any special enquiries regarding the SN110plus.

Volume: 108 l