Eppendorf Research plus single-channel pipette set, pack of 3, 2–20/20–200/100–1000 µL, 3123000918
€684.00
- Pack of 3 with volume ranges of 2–20 µL, 20–200 µL and 100–1,000 µL
- Mechanical air-cushion principle with spring-loaded tip cone
- PerfectPiston made from Fortron, with a quickly detachable base
- Autoclavable at 121 °C for 20 minutes
- Accuracy tested in accordance with ISO 8655
- Includes epT.I.P.S. boxes
Product information "Eppendorf Research plus single-channel pipette set, pack of 3, 2–20/20–200/100–1000 µL, 3123000918"
The Research-plus 3-pack Option 2 comprises three mechanical single-channel pipettes with consecutive volume ranges: 2 to 20 µL, 20 to 200 µL and 100 to 1,000 µL. For a laboratory requiring the same basic equipment for pipetting volumes ranging from microlitres to millilitres, this set covers the three most common routine volumes in a single order.
Three volume ranges, one pipetting principle
All three pipettes in the set operate on the same mechanical air-cushion principle as the Research-plus series, with a spring-loaded tip cone up to 1 mL and a Fortron PerfectPiston in the quick-release lower section. Anyone already familiar with one of the three pipettes in the set will find the other two easy to use straight away.
Each pipette in detail
The 2 to 20 µL pipette (increment 0.02 µL, epT.I.P.S. yellow 53 mm) achieves a systematic measurement error of 1.0 per cent or 0.2 µL at 20 µL and a random measurement error of 0.3 per cent or 0.06 µL. The 20 to 200 µL pipette (increment 0.2 µL, epT.I.P.S. yellow 53 mm) achieves a systematic measurement error of 0.6 per cent (or 1.2 µL) and a random measurement error of 0.2 per cent (or 0.4 µL) at 200 µL. The 100 to 1,000 µL pipette (1 µL increment, epT.I.P.S. blue 71 mm) achieves a systematic measurement error of 0.6 per cent, or 6.0 µL, and a random measurement error of 0.2 per cent, or 2.0 µL, at 1,000 µL. Tests are carried out in accordance with ISO 8655, with ten determinations per volume, and using water in accordance with ISO 3696.
Autoclavable for daily use
The upper and lower parts can be autoclaved together at 121 °C and 1 bar overpressure for 20 minutes; the flasks do not need to be re-greased afterwards. The filter sleeve and protective filter are autoclaved separately.
Selection guide: for which laboratory equipment this set is intended
This set is suitable for workstations that switch between small volumes in the double-digit microlitre range and larger volumes in the millilitre range during routine operations, for example when preparing dilution series or preparing reagents of different concentrations. Those who, on the other hand, work predominantly in the sub-microlitre range or in the millilitre range above 1 mL will find more suitable individual volumes in the Reference 2 or Xplorer ranges, which feature a different operating logic.
Product range overview: Research plus, Reference 2 and Xplorer
Research plus, Reference 2 and Xplorer are all piston-driven pipettes based on the air-cushion principle and each dispense a single volume using an epT.I.P.S. tip. The difference lies in their operation: Research plus is the basic mechanical model with a selection dial and, according to the manufacturer, the lowest weight and lowest operating forces in the Eppendorf pipette family. Reference 2 is operated via a single button and features a four-digit display with a magnifying window for reading volumes from any angle. Xplorer is the electronic version with a selection dial, rocker switch and display, which also allows volumes to be locked and batch runs to be programmed. Anyone who manually sets many different volumes in a single day and values a particularly light weight in the hand will find the Research plus well suited to their needs. For repetitive dispensing of the same volume into many vessels from a single sample, however, a Multipette (M4, E3 or E3x) is the more suitable choice, as it operates on the direct displacement principle and does not re-aspirate after each dispense.
Materials and environmental conditions
The upper section is made of modified polypropylene (PP), polycarbonate (PC) and polyetherimide (PEI), whilst the lower section is additionally made of PVDF, PPS, PEEK, PTFE, EPDM, silicone and stainless steel. The pipettes are operated at 5 to 40 °C and 10 to 95 per cent relative humidity, and are stored without transport packaging at minus 5 to 45 °C. According to the manufacturer, all volumes up to 1 mL feature a spring-loaded tip cone, which reduces the force required to attach and eject the tip during tip changes.
Typical applications
- Sample preparation with varying volumes ranging from a few microlitres to one millilitre
- Preparation of dilution series across several orders of magnitude
- Basic equipment for new workstations where several volume ranges are required simultaneously
- Routine applications in sample preparation and reagent addition
Technical details
| Pipettes included | Research plus variable, 2–20 µL / 20–200 µL / 100–1,000 µL |
| Tips | epT.I.P.S. yellow 53 mm (2–20 µL and 20–200 µL), epT.I.P.S. blue 71 mm (100–1,000 µL) |
| Increment | 0.02 µL (2–20 µL), 0.2 µL (20–200 µL), 1 µL (100–1,000 µL) |
| Systematic measurement error 2–20 µL at 20 µL | 1.0% / 0.2 µL |
| Systematic measurement error 20–200 µL at 200 µL | 0.6 per cent / 1.2 µL |
| Systematic measurement error 100–1,000 µL at 1,000 µL | 0.6% / 6.0 µL |
| Random measurement error 2–20 µL at 20 µL | 0.3% / 0.06 µL |
| Random measurement error 20–200 µL at 200 µL | 0.2% / 0.4 µL |
| Random measurement deviation 100–1,000 µL at 1,000 µL | 0.2% / 2.0 µL |
| Autoclavability | 121 °C, 1 bar gauge pressure, 20 minutes, assembled |
| Operating environmental conditions | 5 to 40 °C, 10 to 95 % relative humidity |
Contents
- 3 x Research plus single-channel pipettes, variable (2–20 µL, 20–200 µL, 100–1,000 µL)
- epT.I.P.S. boxes
Frequently asked questions
- Are all three pipettes in the set autoclavable? Yes, the upper and lower parts can be autoclaved together at 121 °C, 1 bar overpressure, for 20 minutes; the filter sleeve and protective filter are autoclaved separately.
- Do I need to re-lubricate the pistons after autoclaving? No, according to the operating instructions, re-lubrication of the pistons is not necessary; a gravimetric check is recommended.
- What kind of day-to-day laboratory work is this set designed for? For workstations that regularly switch between microlitre and millilitre volumes, such as when carrying out dilution series or sample preparation, without having to purchase a separate pipette for each application.
- Do the tips of the three pipettes fit together? No, each pipette in the set uses its own epT.I.P.S. size to match its volume range; the tips are not designed to be mixed and matched between the three models.
Individual requirements & procurement service
You can enquire about suitable epT.I.P.S. refill packs or spare parts for the Research-plus series via LT Laborhandel’s procurement service. Contact us at lt-laborhandel.de/service.
| Number of channels: | 1-channel |
|---|---|
| Operation: | mechanical |
| Volume setting: | variable |