Skip to main content Skip to search Skip to main navigation

IKA ROTAVISC lo-vi Complete Rotational Viscometer Set, low-viscosity media, TFT touchscreen, 0010013826

Product information "IKA ROTAVISC lo-vi Complete Rotational Viscometer Set, low-viscosity media, TFT touchscreen, 0010013826"

Anyone who routinely tests low-viscosity liquids such as oils, paints, drinks or aqueous solutions needs a rotational viscometer that is ready to measure straight out of the box, rather than having to search for accessories first. The ROTAVISC lo-vi Complete from IKA comes as a fully integrated set comprising the measuring head, stand, spindle set and temperature probe; with its soft measuring spring, it precisely covers the range in which many laboratory quality controls are carried out: from 1 to 6,000,000 mPas.

Rotational principle with a soft measuring spring for low-viscosity media

The lo-vi measuring system uses the softest spring in the ROTAVISC series (spring torque 0.0673 mNm) and is therefore designed for low-viscosity to medium-viscosity samples. The viscosity measurement range extends from 1 to 6,000,000 mPas with an accuracy of 1 per cent of full scale (FSR) and a repeatability of 0.2 per cent FSR. The rotational speed can be adjusted from 0.01 to 200 rpm in increments of 0.01 rpm, which allows even very flat viscosity curves to be resolved with high precision.

TFT touchscreen display with trend graph and timer

The ROTAVISC lo-vi Complete is operated via a TFT touchscreen display, on which speed, torque and temperature are shown simultaneously as setpoint and actual values. An integrated graphing function displays the viscosity curve directly on the instrument, which is useful for identifying shear-thinning or shear-thickening at a glance, without first having to export the measured values to external software. For long measurement series, a timer ranging from 0.017 to 6,000 minutes is available, either in timer mode or continuous operation. A calibration function directly on the device ensures measurement accuracy throughout its service life.

Temperature-controlled measurement with PT 100

As viscosity is highly temperature-dependent, a PT 100.8 temperature sensor is included as standard. It has a resolution of 0.1 Kelvin and, together with the instrument, covers an operating temperature range from minus 100 to 300 degrees Celsius. Users wishing to actively control the temperature of the sample during measurement can utilise an external cooling/heating circulation bath from the range of accessories, such as the HRC 2 lite or the CBC VISC lite, which use the climate-friendly refrigerant R290. Separate measurement chambers are also available as accessories for sealed or particularly small sample volumes, enabling temperature-sensitive or volatile media to be measured under controlled conditions. An RS-232 interface and a micro-USB port are provided for connection to a process control system or the Labworldsoft 6 laboratory equipment software, allowing measurement series to be documented and analysed even outside the TFT display.

Typical applications

With its wide measurement range, designed for low to medium viscosities, the ROTAVISC lo-vi Complete is technically suitable for:

  • Quality control of low-viscosity products such as oils, paints, dip coatings or beverages
  • Incoming goods inspection of raw materials with defined viscosity specifications
  • Recipe and formulation development in research and development
  • In-process spot-check measurements on low- to medium-viscosity batches
  • Training and teaching in rheology, thanks to the clear TFT user interface

Selection guide within the ROTAVISC range

IKA offers the ROTAVISC series in three measurement range variants, which differ primarily in terms of the spring stiffness of the measuring head and the spindle set supplied. All three share the same display, speed range, interfaces and housing:

VariantViscosity measurement rangeSpring torqueStand
ROTAVISC lo-vi Complete, this unit1 to 6,000,000 mPas0.0673 mNmRotastand, manual
ROTAVISC me-vi HELI Complete100 to 40,000,000 mPas0.7187 mNmHelistand, motorised
ROTAVISC hi-vi II Complete800 to 320,000,000 mPas5.7496 mNmRotastand, manual

For most low- to medium-viscosity standard samples in the laboratory, the lo-vi version is the right choice. If the viscosity rises into the range of pastes and highly viscous substances, the ROTAVISC hi-vi II Complete is the more suitable option. If controlled up-and-down movement of the spindle is also required, for example with non-flowing or structurally viscous samples, the ROTAVISC me-vi HELI Complete, with its motorised helix stand, is the ideal addition.

Distinction from the CBC series with integrated temperature control

In addition to the Complete set described here, in which temperature control is provided by an optional external recirculating bath, IKA offers the CBC series – a variant in which a cooling/heating bath is already permanently integrated into the viscometer. For the low-viscosity measurement range, this corresponds to the ROTAVISC CBC lo-vi Package. Users who regularly carry out measurements at defined, actively controlled temperatures thus avoid the need to set up a separate external bath; those who only occasionally work under temperature-controlled conditions will generally find the Complete Set, combined with an external thermostat such as the CBC VISC lite (which can be added as required), to be the more cost-effective option.

Technical details

Viscosity measurement range1 to 6,000,000 mPas
Accuracy (FSR)1 %
Repeatability (FSR)0.2 %
Spring torque0.0673 mNm
Guide raillo-vi
Measuring spindlesSP set-1 (SP 1 to SP 4)
Motor power, output4.8 W
Motor direction of rotationclockwise, with overload protection
DisplayTFT touchscreen with graph function
Speed range0.01 to 200 rpm, setting accuracy ±0.01 rpm
Torque display / measurementYes
Timer0.017 to 6,000 min, timer and continuous operation
Operating temperature-100 to 300 °C, temperature measurement resolution 0.1 K
Connection for external temperature sensorPT 100
Permissible density9999 kg/dm³
Base vessel volume600 ml
Test rigRotary stand, max. stroke 61 mm, Ø 16 mm, dynamic load 5 kg
Stroke of telescopic stand200 mm
InterfacesRS 232, USB (Micro-USB)
Protection classIP 40 in accordance with DIN EN 60529
Dimensions (W x H x D)351 x 629 x 372 mm
Weight7.1 kg
Power rating100 to 240 V, 50/60 Hz, 24 W

Contents

  • 1 x ROTAVISC lo-vi control S000 (measuring head)
  • 1 x Rotastand (tripod)
  • 1 x SP set-1 (standard spindle set SP 1 to SP 4)
  • 1 x PT 100.8 (temperature sensor)
  • 1 x Guard rail (protective rail)

Use, installation, maintenance and disposal

The device is protected against solid foreign objects from 1 millimetre in size in accordance with protection class IP 40; direct contact with splashing water must be avoided. To ensure reproducible measurement results, the stand should be placed on a level work surface with minimal vibration; the guard rail protects the spindle and measuring head from accidental contact during operation. The integrated calibration function allows for regular checks using standard silicone oils from the range of accessories; it is advisable to switch off the power before changing spindles or vessels. Spindles and the measuring vessel should be cleaned thoroughly after each measurement to ensure that dried-on residues do not distort the next measurement. At the end of its service life, the ROTAVISC lo-vi Complete, as an electrical appliance, must not be disposed of with household waste but must be placed in the separate collection for waste electrical and electronic equipment.

Individual requirements & procurement service

Additional spindles and adaptors, measuring chambers, standard silicone oils for calibration, external circulating thermostats and the Labworldsoft 6 laboratory equipment software are available for the ROTAVISC lo-vi Complete as required. Certificates for qualification (IQ/OQ) or factory calibration can also be ordered. Through our procurement service, LT Laborhandel assists you in putting together the right accessories and selecting from our range of viscometers.

Frequently Asked Questions

Is the ROTAVISC lo-vi Complete ready for immediate use?
Yes, the measuring head, rotary stand, SP set-1 spindle set, PT 100.8 temperature sensor and protective bracket are all included in the scope of delivery.

What viscosity range is the instrument designed for?
For 1 to 6,000,000 mPas, using the softest measuring spring in the ROTAVISC range for low- to medium-viscosity media.

Can the sample be temperature-controlled during the measurement?
Yes, using the supplied PT 100 probe and an optional external circulating bath from the range of accessories, such as the CBC VISC lite.

What is the difference between the lo-vi and hi-vi II models?
The hi-vi II Complete measures up to 320,000,000 mPas using a stiffer spring and is therefore designed for pastes and highly viscous media.

Can I calibrate the device myself?
A calibration function is built in, and suitable standard silicone oils are available as accessories.

What interfaces are available for data transfer?
RS 232 and Micro-USB, compatible with the Labworldsoft 6 laboratory instrument software.

Skip product gallery

Matching accessories

Discount
IKA CBC VISC lite Kälte- und Wärme-Bad-Umwälzthermostat, -25 bis 125 °C, für ROTAVISC Viskosimeter
IKA CBC VISC lite Refrigerated and Heating Bath Circulator, -25 to 125 °C, for ROTAVISC Viscometers, 20102007
Herstellernummer: 0020102007

Rotational viscometers only deliver reliable measured values if the sample stays precisely temperature-controlled during the measurement. The IKA CBC VISC lite (0020102007) is a refrigerated and heating bath circulator designed specifically for combination with IKA ROTAVISC rotational viscometers, with a working temperature range of -25 °C to 125 °C and space for two standard beakers in the open bath. While one sample is being measured in the bath, a second one can already be pre-conditioned, which saves waiting time in measurement series with several samples. Temperature range and control accuracy for viscosity measurement The heating power of 2,000 watts and a cooling capacity of 450 watts at 20 °C to 120 watts at -20 °C cover the working temperature range of -25 °C to 125 °C. The temperature stability to DIN 12876 is ±0.05 kelvin both at 70 °C and at -10 °C, controlled by a PID controller with PT1000 working and safety probes. This stability is important in viscosity determination, because the viscosity of many media depends strongly on temperature and even small temperature fluctuations can falsify the measured values. Environmentally friendly refrigerant and optimised water flow The natural, environmentally friendly R290 is used as refrigerant, with a fill quantity of 48 grams, designed for a maximum pressure of 21 bar. Optimised water flow management in the insulated bath ensures fast and precise temperature control. The noise level is 53 dB(A), and an air filter that is easy to remove and clean keeps the cooling performance constant over time. External pump connection for standard vessels to DIN Via the external pump connection, the CBC VISC lite can also be used to heat standard vessels to DIN for absolute viscosity determination with ELVAS spindles or for double-walled beakers. The pressure/suction pump reaches a maximum delivery pressure of 0.35 bar, a suction pressure of 0.15 bar and a flow rate of up to 18 litres per minute via the M16x1 connection. The fill volume of the bath is 5.5 to 8.5 litres, the bath opening measures 150 x 218 millimetres and the total height of the bath is 234 millimetres. The CBC VISC lite can thus be used flexibly both for direct measurement in the open bath and for temperature control of an external, closed test vessel, depending on which specific measurement method the respective viscosity determination requires in each case. Compact design for the workstation at the viscometer With external dimensions of 414 x 392 x 505 millimetres and a weight of 29 kilograms, the CBC VISC lite is considerably more compact than the CBC 5 series and is designed for installation directly next to a ROTAVISC viscometer. The IP 21 protection class and the permissible ambient temperature range of 5 to 32 °C at up to 80 % relative humidity cover the usual laboratory conditions. The device can be connected to a PC via the USB and RS 232 interfaces. Safety classification and insulation of the bath To DIN 12876, the CBC VISC lite is classified in safety class III with the marking FL, and the adjustable safety circuit ranges from 0 to 135 °C. The insulated bath reduces heat loss to the outside and, together with the optimised water flow management, helps set temperatures to remain stable even over longer measurement series, without the ambient temperature in the laboratory influencing the measurement. For measurement series with frequently changing temperature steps, this combination of fast control and stable insulation is a practical advantage over open, uninsulated water baths. Typical applications As a bath circulator tailored specifically to viscosity measurement, the CBC VISC lite is used above all where temperature stability and sample throughput have to work together: Temperature control of samples for rotational viscometers of the ROTAVISC series directly in the open bath Parallel pre-conditioning of a second sample while the first is already being measured Absolute viscosity determination with ELVAS spindles in standard vessels to DIN via the external pump connection Temperature-dependent viscosity series over a wide range from -25 °C to 125 °C Quality control and research in the coatings, paint and food industries, where viscosity is recorded as a test parameter Technical details Device typeRefrigerated and heating bath circulator Class to DIN 12876III (FL) RefrigerantR290, 48 g, max. 21 bar Heating power2,000 W Cooling capacity at 20 °C / 10 °C / 0 °C / -10 °C / -20 °C450 W / 420 W / 380 W / 250 W / 120 W Working temperature range-25 °C to 125 °C Temperature stability to DIN 12876 (at 70 °C and -10 °C)±0.05 K Temperature controlPID controller, PT1000 working and safety probes DisplayLED Adjustable safety circuit0 to 135 °C Bath fill volume5.5 to 8.5 l Bath opening (W x D) / total bath height150 x 218 mm / 234 mm Pump typePressure/suction pump Max. pump pressure / max. suction pressure0.35 bar / 0.15 bar Max. flow rate18 l/min Pump connectionM16x1 Noise level53 dB(A) Dimensions (W x H x D)414 x 392 x 505 mm Weight29 kg Protection class to DIN EN 60529IP 21 Permissible ambient temperature / rel. humidity5 to 32 °C / 80 % InterfacesUSB, RS 232 Power supply230 V, 50/60 Hz Power consumption2,320 W Scope of delivery 1 x IKA CBC VISC lite Refrigerated and Heating Bath Circulator Individual requirements and procurement service To connect the CBC VISC lite to ROTAVISC viscometers or to external standard vessels, we support you through our procurement service in selecting suitable ELVAS spindles, hose sets and heat transfer fluids. We are also at your side with spare parts and individual calibration. Further information and contact options can be found at lt-laborhandel.com/service.

Sale price: €2,913.00 Manufacturer's RRP: €3,884.00 25% below manufacturer's RRP
Discount
IKA HRC 2 lite Circulator, Cooling/Heating -10 to 100 °C, R290, 0020104311
IKA HRC 2 lite Circulator, Cooling/Heating -10 to 100 °C, R290, 0020104311
Herstellernummer: 0020104311

Some applications need both cooling and heating in the same series of experiments, for example when a bioreactor first has to be cooled down to 5 degrees Celsius and then heated to 80 degrees Celsius. The IKA HRC 2 lite covers exactly this case: as a refrigerated and heating circulator it works in the range from minus 10 to 100 degrees Celsius, with the natural refrigerant R290 and a heating power of 1000 watts in a compact benchtop unit. One unit for cooling and heating from minus 10 to 100 degrees Celsius The cooling capacity is 400 watts at 20 degrees Celsius and falls with decreasing target temperature to 350 watts at 10 degrees, 250 watts at 0 degrees and 100 watts at minus 10 degrees Celsius. For the opposite direction, a heating power of 1000 watts is available. The temperature stability to DIN 12876 is ±0.1 kelvin in the colder range around minus 10 degrees Celsius and improves to ±0.05 kelvin at plus 70 degrees Celsius. Control is by PID controller with a PT1000 working temperature sensor and a separate PT1000 safety temperature sensor, and the LED display provides a resolution of 0.1 kelvin with the same setting accuracy. Safety functions and refrigerant R290 The HRC 2 lite has visual and acoustic warning functions as well as its own over-temperature warning, and the safety circuit is adjustable between 0 and 110 degrees Celsius. To DIN 12876 the unit is classified as Class II with the designation FL, which makes it suitable for operation with flammable liquids as long as there is a sufficient safety margin between the flash point and the set maximum temperature. The refrigerant used is R290, with a fill of 79 grams at a maximum permissible pressure of 21 bar. Practical note: R290 (propane) is a natural but flammable hydrocarbon refrigerant, so, as is usual for all laboratory equipment with this type of refrigerant, we recommend a sufficiently ventilated installation location without open ignition sources in the immediate vicinity of the unit. HRC 2 lite or HRC 2 control: choosing the right specification level The HRC 2 lite and the IKA HRC 2 control share the operating principle of a refrigerated and heating circulator, but differ clearly in equipment and specifications: the HRC 2 lite uses the natural refrigerant R290 instead of the synthetic R134a of the HRC 2 control, is classified as Class II instead of Class III to DIN 12876 and delivers 1000 instead of 1500 watts of heating power, with a somewhat narrower working temperature range of minus 10 to 100 degrees Celsius instead of minus 30 to 100 degrees Celsius. In addition, the HRC 2 lite does without the analogue output of the HRC 2 control and is limited to RS232 and USB-B. The HRC 2 control itself has been discontinued since 31 July 2026, and its successor is the IKA HRC 2 GREEN control with the natural refrigerant R290. If cooling only is needed, without a heating function, the dedicated recirculating chiller IKA RC 2 lite is the leaner and technically closely related alternative, with the same housing and the same pump but without a heating element. If, conversely, only heating is needed, without active cooling, a simple IKA HB digital heating bath is often sufficient. Typical applications As a refrigerated and heating circulator with a wide temperature range, the HRC 2 lite is suitable, among other things, for: Temperature control of laboratory reactors and bioreactors over changing temperature cycles from cold to warm Temperature control of open baths for sample preparation or reaction control Temperature control source for connected measuring instruments that need a stable reference temperature Use as a peripheral device with device-to-device communication via RS232 or USB to main units, as well as control and monitoring via laboratory software Connection, installation and ongoing operation The integrated pressure/suction pump reaches a maximum flow rate of 18 litres per minute at a maximum pump pressure of 0.35 bar on the pressure side and 0.15 bar on the suction side, connected via M16x1 threaded connections. The scope of delivery includes two olives for DN 8 hose connections and a USB-A-B cable for direct PC connection. The filling volume of the internal bath is between 1 and 3.5 litres, and a calibration option is available to check the temperature display regularly against a reference sensor. For continuous operation the HRC 2 lite is designed for a permissible duty cycle of 100 percent and remains comparatively quiet at 51 dB(A); the permissible ambient temperature range is 5 to 32 degrees Celsius at a maximum of 80 percent relative humidity. One unit instead of two: when the dual function pays off Anyone who regularly has to run both cooling and heating cycles in the same experimental set-up saves one unit space and one additional hose connection with the HRC 2 lite, compared with a combination of a separate recirculating chiller and heating bath. For applications that only ever cool or only ever heat, on the other hand, the respective specialised single unit is often the more economical choice, as the heating power of 1000 watts and the cooling capacity steps down to 100 watts at minus 10 degrees Celsius represent compromises compared with the respective specialised units. As for the pump design, the same applies as for all units in this series: the maximum flow rate of 18 l/min is only reached with short hose runs without significant back pressure, and longer lines or several connected consumers reduce the flow that actually arrives in line with the increasing line resistance. Technical details Device typeRefrigerated and heating circulator Classification to DIN 12876Class II, designation FL Working temperature range-10 to 100 °C Temperature stability (DIN 12876, -10 °C / +70 °C)±0.1 K / ±0.05 K Temperature sensors (working / safety)PT1000 / PT1000 Display resolution / setting accuracy0.1 K Heating power1000 W Cooling capacity at 20 / 10 / 0 / -10 °C400 / 350 / 250 / 100 W RefrigerantR290, 79 g, max. 21 bar Warning functionsvisual, acoustic, over-temperature warning Adjustable safety circuit0 to 110 °C Pump typePressure/suction pump Max. pump pressure (pressure / suction side)0.35 bar / 0.15 bar Max. flow rate (0 bar back pressure)18 l/min Pump connectionsM16x1 Filling volume1 to 3.5 l Calibration optionyes Permissible duty cycle100 % Noise level51 dB(A) InterfacesRS232, USB-B Protection class to DIN EN 60529IP 21 Permissible ambient temperature5 to 32 °C Permissible relative humiditymax. 80 % Dimensions (W x H x D)225 x 385 x 430 mm Weight24.9 kg Voltage / frequency / connected load230 V, 50/60 Hz, 1250 W (other variants 115 V / 100 V available) Scope of delivery 1 x IKA HRC 2 lite refrigerated and heating circulator (ident no. 0020104311) 2 x olive for DN 8 hoses 1 x USB-A-B cable Regulatory information As an electrical appliance, the HRC 2 lite is subject to the German Electrical and Electronic Equipment Act (ElektroG) and is registered with Stiftung EAR under the WEEE registration number DE 66784213. At the end of its service life the unit must not be disposed of with household waste but must be taken to a separate collection for waste electrical equipment. The refrigerant R290 contained in the refrigerant circuit must be removed by an approved refrigerant recovery company before disposal. Frequently asked questions about the HRC 2 lite What is the difference between the HRC 2 lite and the HRC 2 control?The HRC 2 lite uses R290 instead of R134a, is classified as Class II instead of Class III to DIN 12876, delivers 1000 instead of 1500 watts of heating power, covers a somewhat narrower working range of minus 10 to 100 degrees Celsius and does without the analogue output of the HRC 2 control. The HRC 2 control has been discontinued since 31 July 2026, and its successor is the IKA HRC 2 GREEN control. Can the HRC 2 lite also be used as a pure chiller without a heating function?Technically yes, the heating function can simply be left unused. If only cooling is needed permanently, the identical-design IKA RC 2 lite without a heating element is usually the more economical choice. Which hoses fit the HRC 2 lite?Connection is via M16x1 thread, and two olives for DN 8 hose connections are supplied with the unit to match. Do I need to observe anything special with the refrigerant R290?Yes, R290 (propane) is flammable, even though it counts as a natural refrigerant. A well-ventilated installation location without ignition sources in the immediate vicinity of the unit is therefore usual and sensible. Why is the DIN 12876 designation FL important?It indicates that the unit is also suitable for operation with flammable liquids as the bath medium, provided their flash point maintains a sufficient safety margin to the set maximum temperature. Can the HRC 2 lite be calibrated?Yes, the unit has a calibration option to check the temperature display regularly against a reference sensor. How loud is the HRC 2 lite in continuous operation?At 51 dB(A) the unit remains pleasantly quiet even in continuous operation at a permissible duty cycle of 100 percent. Individual requirements & procurement service For suitable connection hoses, additional valves, adapters or connection to existing laboratory and bioreactors, our procurement service will support you. If you have questions about certificates and certifications for the HRC 2 lite, please do not hesitate to contact us.

Sale price: €2,352.00 Manufacturer's RRP: €3,136.00 25% below manufacturer's RRP