Memmert HCP105 humidity chamber, 107 l, +18 °C to +90 °C
€7,495.00
- Active humidification and dehumidification (20–95 % rh) for extremely stable climatic conditions.
- Integrated sterilisation programme at +160 °C to prevent cross-contamination.
- Optimal size with a 107-litre capacity for versatile stability testing.
- TwinDISPLAY ControlCOCKPIT for intuitive programming and GLP documentation.
- Electropolished stainless steel interior guarantees maximum hygiene and durability.
- Internal data logger (10-year capacity) for seamless monitoring of your processes.
- Manufacturer’s part number (MPN): HCP105
Product information "Memmert HCP105 humidity chamber, 107 l, +18 °C to +90 °C"
Precise environmental simulation for demanding stability tests
The Memmert HCP105 humidity chamber is the ideal high-performance unit for applications requiring precise control of temperature and humidity. With an internal chamber volume of 107 litres, this model offers the perfect balance between capacity and space requirements in the laboratory. It is specially designed for stability testing, environmental simulations or accelerated life cycle testing in the pharmaceutical industry, electronics development or materials research. Memmert’s HCP series guarantees unrivalled homogeneity within the chamber, ensuring that every sample is exposed to identical conditions – a must for reproducible results.
Active humidity control for reproducible results
A key feature of the HCP105 is its active humidification and dehumidification system. Unlike passive systems, active control enables an extremely wide humidity range of 20% to 95% RH, which is precisely regulated regardless of the temperature. Humidity is supplied via an external water tank and a self-priming pump, thereby minimising the risk of condensation forming inside the chamber. This is particularly critical during long-term tests, where even the slightest fluctuations could compromise the validity of the results. The digital control system guarantees an extremely fast recovery time after the door is opened, ensuring that your process is barely interrupted.
Maximum hygiene thanks to an integrated sterilisation programme
The interior of the HCP105 is made from high-quality, electropolished stainless steel (grade 1.4301), which not only guarantees a long service life but also enables easy, residue-free cleaning. A key safety feature is the integrated sterilisation programme: the entire interior, including all sensors, can be sterilised at +160 °C. This reliably prevents cross-contamination and saves time during preparation between different test series. The chamber is controlled via the intuitive TwinDISPLAY system, which clearly displays all parameters on two high-resolution screens.
Intelligent control and comprehensive documentation
In regulated industries, the documentation of climate data is essential. The HCP105 features an internal data logger with a capacity of at least 10 years, which records all measured values in a tamper-proof manner. Via the Ethernet interface or the USB port, this data can be exported at any time and analysed using the AtmoCONTROL software. Multiple temperature and humidity safeguards, such as the electronic temperature monitor (TWW) and the AutoSAFETY system, protect your valuable samples from critical deviations around the clock.
Technical details
- Internal volume: 107 litres
- Operating temperature range: +18 °C to +90 °C (with humidity)
- Temperature setting range: +18 °C to +160 °C (without humidity / sterilisation)
- Temperature control accuracy: 0.1 °C
- Humidity range: 20% to 95% rh
- Humidity control accuracy: 0.5% rh
- Internal dimensions (W x H x D): 560 x 480 x 400 mm
- External dimensions (W x H x D): 745 x 947 x 591 mm (excluding door handle and connections)
- Interior material: Stainless steel 1.4301 (ASTM 304), electropolished finish
- Casing material: Textured stainless steel, galvanised rear panel
- Max. number of shelves: 5 (2 wire shelves included)
- Max. load per unit: 120 kg (20 kg per wire rack)
- Sterilisation programme: Fixed at +160 °C (duration approx. 4 hours)
- Interfaces: Ethernet (LAN), USB
- Data logger: Internal memory for log data (capacity for at least 10 years)
- Safety: Electronic temperature monitoring (TWW), protection class 3.1
- Mains connection: 230 V, 50/60 Hz; power consumption approx. 1500 W
- Weight: approx. 75 kg net
Scope of delivery
- HCP105 humidity chamber
- 2 stainless steel grates
- Factory calibration certificate (measuring point: +60 °C)
- Water connection kit including hose
- USB stick with AtmoCONTROL software and user manual
Accessories & Advice
Boost your research with the right setup: whether it’s additional stainless steel grating for maximum capacity utilisation, specialised feed-throughs for external sensors, or professional IQ/OQ qualification carried out directly at your premises – we’ll configure your HCP105 precisely to your validation specifications. Are you planning to integrate the system into your Laboratory Information Management System (LIMS) or do you need advice on standby operation? Take advantage of our dynamic service for personalised expert advice. Our team will be happy to provide you with a bespoke quote that will sustainably enhance your process reliability.
| Volume: | 107 l |
|---|