Skip to main content Skip to search Skip to main navigation

Heidolph Titramax 1000 Bench-top Shaker, 6 Microtitre Plates, with Incubation Function, 544-12200-00

Product information "Heidolph Titramax 1000 Bench-top Shaker, 6 Microtitre Plates, with Incubation Function, 544-12200-00"

Anyone who regularly handles six microtitre plates at the same time or also needs to temperature-control samples will find the smaller models in the Titramax range fall short. The Titramax 1000 retains the orbit and speed range of the Titramax 100, but increases the usable area to 290 x 258 millimetres and can be upgraded to a temperature-controlled system with the Incubator 1000 heating module.

Same orbit, larger surface area

The 1.5-millimetre circular vibrating orbit and the speed range of 150 to 1,350 revolutions per minute are the same as on the Titramax 100, but the working area increases from 220 x 220 to 290 x 258 millimetres. This means there is space for up to six microtitre plates at the same time, and the maximum permissible load increases to 5 kilograms. A rubber mat with recesses and spacing holds the plates securely and allows for convenient loading and unloading.

Expandable to an incubation system

The accessories section on the manufacturer’s website lists a heating module for the Titramax 1000, as well as a low and a high incubation hood. When combined with the Titramax 1000 heating module and an incubation hood, samples can be temperature-controlled whilst being shaken, for example, for applications requiring both agitation and heat simultaneously. Anyone wishing to purchase this combination as a ready-configured package will find it in the Titramax package. The smaller Titramax 100 does not have a link to the same accessories section; temperature control is therefore a unique selling point of the Titramax 1000 within the range.

Typical applications

  • Simultaneous processing of up to six microtitre plates
  • Shaking with simultaneous temperature control in combination with a heating module and an incubation hood
  • Higher sample throughputs in batch experiments
  • Laboratory routines requiring more space per run

Drive and operating data

The shaded-pole motor draws 25 watts and keeps the sound pressure level below 50 dB(A). A timer limits runs to a maximum of 120 minutes with an audible signal at the end; for longer processes, the unit continues to run in continuous operation. Weighing 8 kilograms and measuring 320 x 128 x 375 millimetres, the Titramax 1000 requires more floor space than the smaller models in the range. The same ambient temperature limits apply as for the smaller models: 5 to 31 degrees Celsius at up to 80 per cent relative humidity; at higher temperatures up to 40 degrees Celsius, the permissible humidity decreases linearly to 50 per cent. The unit is designed for installation at altitudes of up to 2,000 metres above sea level, as well as for overvoltage category II and pollution degree 2; it compensates for mains voltage fluctuations of up to 10 per cent without any loss of function.

Technical details

Dimensions (W x H x D)320 x 128 x 375 mm
Weight8 kg
Maximum permissible load5 kg
DriveSlotted-pole motor
Speed range150 to 1,350 rpm
Orbit1.5 mm
Type of movementcircular vibration
Working area290 x 258 mm
Speed controlanalogue
Sound pressure level< 50 dB(A)
Power consumption25 W
Unit fuseM 1.25 A (230 V) or T 2.0 A (115 V)
Permissible mains voltage fluctuations±10 %
Overvoltage categoryII
Pollution degree2
Operating temperature5 °C to 31 °C at up to 80 % relative humidity, 32 °C to 40 °C at up to 50 % relative humidity (linearly decreasing)
Installation altitudeup to 2,000 metres above sea level
Overheating protectionself-resetting

Scope of delivery

  • 1x Titramax 1000 platform shaker, order number 544-12200-00
  • Mains cable for EU mains connection

The heating module and incubation hoods are not included with this order number and must be ordered separately if required; they are supplied as a complete set in the Titramax package.

Regulatory information

The device is an electrical appliance within the meaning of the German Electrical and Electronic Equipment Act (ElektroG). The WEEE registration number of the manufacturer, Heidolph Scientific Products GmbH, is DE 50705753. Information on the return of end-of-life equipment can be found on our Environment and Disposal page.

Warranty and procurement service

Heidolph Scientific Products GmbH, Schwabach, provides a 3-year manufacturer’s warranty against defects in materials and workmanship. This excludes glass and wear parts, damage caused during transport and damage resulting from improper use. If the appliance is registered with Heidolph, the warranty begins on the date of purchase; otherwise, it begins on the date of manufacture as indicated by the serial number. Your statutory warranty rights remain unaffected. Our procurement service can, if required, assemble a heating module and incubation hood suitable for the Titramax 1000.

Max. speed: 1,350 rpm
Power supply: Mains operation
Supply form: Single unit