Skip to main content Skip to search Skip to main navigation

Memmert Humidity Chamber HCP105, 107 l, +18 to +90 °C, 20 to 95 % rh

Product information "Memmert Humidity Chamber HCP105, 107 l, +18 to +90 °C, 20 to 95 % rh"

The Memmert Humidity Chamber HCP105 comes with two stainless steel perforated sheets from the factory, and up to six shelves are possible. The internal volume is 107 litres, and the interior measures 560 x 480 x 400 mm (W x H x D, depth reduced by 35 mm for the fan). Temperature and relative humidity are actively controlled: +18 to +90 °C and 20 to 95 % rh, also settable with rh-Off. Operation is via the ControlCOCKPIT TwinDISPLAY.

Housing, weight and loading

Externally the HCP105 measures 719 x 850 x 591 mm plus 56 mm door handle, and the net weight is approx. 75 kg. The whole unit may be loaded with 90 kg, and each of the six possible shelves with 15 kg. The interior is made of stainless steel 1.4301 (ASTM 304). Behind the outer door sits a heated inner glass door, and the outer door is fully insulated and has 2-point locking. For supply lines there is a humidity-tight silicone feedthrough with 40 mm inner diameter in the rear wall, centre left. The chamber is transported upright.

Humidity control and water supply

Humidity is supplied by a self-priming pump that draws from an external container. The water canister with connection hose is supplied with the chamber. The germ barrier is implemented as hot steam generation, and dehumidification runs via a sterile filter. The water must be demineralised or fully deionised, with a conductivity of 5 to 10 µS/cm and a pH value between 5 and 7. A central water supply for connection to the domestic water supply is available as an accessory. The setting accuracy of humidity is 0.5 % rh. To avoid condensate formation, door and rear wall are additionally heated, plus all-round heating on six sides.

Programming and documentation

With AtmoCONTROL, programs can be written, managed and transferred to the chamber via Ethernet or USB. The digital countdown clock ranges from 1 minute to 99 days and shows the target time. SetpointWAIT lets the process time run only from the point when the setpoint temperature has been reached. Three values each for temperature and humidity can be freely calibrated. Also settable are Celsius or Fahrenheit, time zone and summer and winter time. If the power fails, the program sequence remains saved.

Fault detection and protection

The two Pt100 sensors (DIN class A, 4-wire measurement) check each other, and in the event of a fault the remaining sensor takes over. AutoSAFETY tracks a freely selectable upper and lower temperature limit to the setpoint. If it is violated, the chamber gives a visual and acoustic alarm, and when exceeded the heating switches off. The self-diagnosis system covers temperature and humidity control. The environment may be 10 to 30 °C warm and at most 70 % rel. humidity, non-condensing, and the installation height is at most 2,000 m above sea level. At the back, 15 cm of wall distance are to be kept, 20 cm at the top, 5 cm at the sides.

HCP105, HCP50 or HCP150

Compared with the HCP50 (56 l, 5 shelves, 75 kg), the HCP105 brings about twice the volume and a sixth shelf. Upwards follows the HCP150 with 156 l, ten shelves and 120 kg total load. The ICH110 climate chamber has almost the same volume at 108 l, but sets -10 to +60 °C and 10 to 80 % rh. Anyone who wants to go up to +90 °C and 95 % rh stays with the HCP.

Typical applications

The HCP series is intended for these applications:

  • Environmental testing
  • Environmental simulations
  • Accelerated lifetime tests
  • 85/85 tests according to IEC 60068-2-67 and IEC 60068-2-78

Technical details

Internal volume107 l
Internal dimensions (W x H x D)560 x 480 x 400 mm (D less 35 mm for fan)
External dimensions (W x H x D)719 x 850 x 591 mm (D plus 56 mm door handle)
Weight netapprox. 75 kg
Shelvesmax. 6, of which 2 perforated sheets as standard
Maximum load90 kg total, 15 kg per shelf
Setting temperature range+18 to +90 °C
Working temperature range7 °C above room temperature to +90 °C
Setting accuracy temperature0.1 °C
Setting range humidity20 to 95 % rh and rh-Off
Setting accuracy humidity0.5 % rh
Sensors2x Pt100 DIN class A, 4-wire measurement
ControlControlCOCKPIT TwinDISPLAY, adaptive PID microprocessor controller
Waterdemineralised or fully deionised, 5 to 10 µS/cm, pH 5 to 7
Electrical connection 230 V50/60 Hz, approx. 1,720 W
InterfacesEthernet LAN, USB
Environment10 to 30 °C, max. 70 % rel., non-condensing
Max. installation height2,000 m above sea level
MarkingsCE, VDE GS, cMETus

Other sizes in the series

The HCP humidity chambers come in four sizes from 56 to 241 litres, each with TwinDISPLAY:

ModelInternal volumeSetting temperatureHumidity range
HCP5056 l+18 to +90 °C20 to 95 % rh
HCP105 (this page)107 l+18 to +90 °C20 to 95 % rh
HCP150156 l+18 to +90 °C20 to 95 % rh
HCP240241 l+18 to +90 °C20 to 95 % rh

Scope of delivery

  • 1x Memmert Humidity Chamber HCP105
  • 2x stainless steel perforated sheet
  • Factory calibration certificate for 60 °C at 75 % rh (measuring room centre of usable space)
  • Heated inner glass door
  • Fully insulated stainless steel door, 2-point locking (compression door closure)
  • Silicone feedthrough with 40 mm inner diameter, rear wall, centre left
  • Water canister including connection hose

Regulatory information

The unit is an electrical device within the meaning of the German Electrical and Electronic Equipment Act (ElektroG). The WEEE registration number of the manufacturer Memmert GmbH + Co. KG is DE 66812464. Notes on returning old devices can be found on our page Environment and disposal.

Individual requirements and procurement service

Factory options for the HCP105, specified when ordering, are: 115 V connection voltage, lateral 23 mm feedthrough left or right, 4-20 mA outputs for the actual value of temperature and humidity, a lockable door, a potential-free fault message contact and MobileALERT for alarm notification by SMS (requires the fault message contact). Accessories listed include, among others, further perforated sheets and grid shelves, the central water supply with or without filter cartridges, USB-Ethernet adapter, Ethernet cable (5 m), USB user ID stick, FDA licence, stacking kit and base frames (130 mm or 622 mm levellable). Calibration at the factory or according to DAkkS, IQ/OQ documents and on-site maintenance can be ordered in addition. Send us your requirements, and we will put together the suitable package.

Volume: 107 l