Memmert Pass-Through Oven UF450TS, 449 l, +20 to +250 °C, TwinDISPLAY

Product information "Memmert Pass-Through Oven UF450TS, 449 l, +20 to +250 °C, TwinDISPLAY"

The Memmert Pass-Through Oven UF450TS offers 449 litres of internal volume and, like the UF750TS, is connected via three-phase current, while the smaller models UF160TS and UF260TS use 230 V or 115 V. Two fully insulated stainless steel doors on the clean and unclean side continue to provide the airlock function between the grey area and the clean room, controlled via TwinDISPLAY in the range +20 to +250 °C.

Design: more load capacity per shelf

The interior measures 1,040 x 720 x 600 mm and the housing externally 1,224 x 1,233 x 782 mm plus the door handles. 2 electropolished stainless steel grid shelves are included as standard, and up to 8 shelves can be fitted. Unlike the smaller TS models, up to 30 kg per shelf is permissible, up to 300 kg in total. The net weight is around 213 kg, and packed in the carton (1,330 x 1,450 x 1,050 mm) around 279 kg.

Control: TwinDISPLAY for larger batches

The ControlCOCKPIT controller with TwinDISPLAY and two TFT colour graphic displays also controls the UF450TS, and HeatBALANCE adjusts the heating power distribution between the upper and lower heating group between -50 and +50 per cent. SetpointWAIT starts the process time only after the setpoint temperature has been reached, and the electronically controllable air turbine distributes the heat segment-specifically in 10 per cent steps. Programmes are managed via Ethernet or USB with AtmoCONTROL, and the sequence remains stored in the event of a power failure. The setting accuracy is 0.1 °C up to a setpoint temperature of 99.9 °C and 0.5 °C from 100 °C.

Electrical connection with three-phase current

The UF450TS is connected at 400 V and 3 x 230 V without neutral conductor at approx. 4,800 W, 50/60 Hz. An installation altitude of up to 2,000 m above sea level is permissible at +5 °C to +40 °C ambient temperature and a maximum of 80 % relative humidity, non-condensing. The cabinet may only be transported upright.

Safety under DIN 12880

Two redundant Pt100 sensors of DIN class A monitor each other in 4-wire measurement, and a mechanical temperature limiter of protection class 1 under DIN 12880 switches off the heating around 20 °C above the maximum cabinet temperature. AutoSAFETY additionally monitors over- and under-temperature with an optical and acoustic alarm, and a self-diagnosis system helps with fault finding.

Choosing: between UF260TS and UF750TS

Anyone who reaches the load limit of 20 kg per shelf with the smaller Memmert Pass-Through Oven UF260TS or needs a three-phase connection will find the suitable upgrade in the UF450TS with 30 kg per shelf and 300 kg total load. For even more capacity at the same load per shelf, the Memmert Pass-Through Oven UF750TS with 749 litres and 14 shelves is the next step. Without an airlock function and with access from one side, the likewise 449 litre Memmert Universal Oven UF450plus is an option, and its temperature range extends to +300 °C.

Typical applications

With a permissible load of 30 kg per shelf, the UF450TS is also suitable for heavier load material that is passed directly between the grey area and the clean room, with a reduced risk of contamination.

  • Airlock transfer of larger or heavier load material and samples between the grey area and the clean room
  • Loading clean room areas with batches of over 20 kg per shelf
  • Drying and tempering of load material with subsequent transfer into the clean room
  • Time saved when loading through direct access from both sides

Technical details

Setting temperature range+20 to +250 °C
Working temperature rangemin. 10 °C above room temperature to +250 °C
Setting accuracyup to 99.9 °C: 0.1 °C, from 100 °C: 0.5 °C
Display resolution, actual value0.1 °C
Temperature sensor2x Pt100, DIN class A, 4-wire measurement, redundant
TimerDigital countdown timer with target time, 1 min to 99 days
Internal volume449 l
Internal dimensions (W x H x D)1,040 x 720 x 600 mm
External dimensions (W x H x D)1,224 x 1,233 x 782 mm (plus 2x56 mm door handle)
Net weightapprox. 213 kg
Gross weight (carton)approx. 279 kg
Dimensions incl. carton1,330 x 1,450 x 1,050 mm
Connection400 V and 3 x 230 V no neutral conductor, 50/60 Hz, approx. 4,800 W
Shelves as standard / maximum2 / 8
Max. load, device300 kg
Max. load per shelf30 kg
ControlControlCOCKPIT TwinDISPLAY, PID microprocessor controller, 2x TFT colour graphic display
Additional functionsHeatBALANCE, SetpointWAIT, AutoSAFETY
Permissible ambient temperature+5 °C to +40 °C
Permissible humiditymax. 80 % rel., non-condensing
Maximum installation altitude2,000 m above sea level
Electrical categoryOvervoltage category II, pollution degree 2
MarkingsCE, EAC

Other sizes in the series

The UF450TS is one of five sizes of the TS series, all with TwinDISPLAY, airlock function and a setting range of +20 to +250 °C.

ModelInternal volumeSetting temperatureDisplay
UF160TS161 l+20 to +250 °CTwinDISPLAY
UF260TS256 l+20 to +250 °CTwinDISPLAY
UF450TS (this page)449 l+20 to +250 °CTwinDISPLAY
UF750TS749 l+20 to +250 °CTwinDISPLAY
UF1250TS (on request)1,248 l+20 to +250 °CTwinDISPLAY

Scope of delivery

  • Pass-through oven UF450TS with fully insulated stainless steel doors on both sides
  • Factory calibration certificate (calibration at +160 °C)
  • 2x stainless steel grid shelf, electropolished
  • Installation on feet

Regulatory information

The device is an electrical device within the meaning of the German Electrical and Electronic Equipment Act (ElektroG). The WEEE registration number of the manufacturer Memmert GmbH + Co. KG is DE 66812464. Information on returning old devices can be found on our page Environment and disposal.

Individual requirements and procurement service

For the UF450TS, options available include a full-view glass door with 4-layer insulating glazing up to 250 °C, a door interlock for alternating opening, an additional Pt100 sensor and a 4-20 mA interface. In addition, reinforced stainless steel grid shelves, slide-in and base drip trays, a built-in frame set for wall installation or a USB-Ethernet adapter can be ordered. For qualification and maintenance, we arrange factory and DAkkS calibration, IQ/OQ documentation under DIN 12880:2007-05, PQ qualification as well as instruction and maintenance contracts. Our procurement service supports you with individual requirements and with selecting the right equipment.