Memmert Pass-through oven UF260TS, 256 l, +20 to +250 °C, TwinDISPLAY

ⓘ  This product description has been translated from German using AI. Despite careful review, we do not guarantee that the translation is error-free; in case of doubt, the German version shall prevail.

Product information "Memmert Pass-through oven UF260TS, 256 l, +20 to +250 °C, TwinDISPLAY"

With 256 litres of internal volume, the Memmert Pass-through oven UF260TS offers more capacity than the most compact model of the TS series, with the same airlock function between the grey area and the clean room. Two fully insulated stainless steel doors on opposite sides pass load material directly between the room areas, controlled via TwinDISPLAY in the range +20 to +250 °C.

Design and dimensions

The interior measures 640 x 800 x 500 mm and the housing externally 825 x 1,314 x 682 mm plus the door handles. 2 electropolished stainless steel grid shelves are included as standard, up to 9 shelves can be fitted, with up to 20 kg per shelf and up to 300 kg total permissible load. The net weight is around 138 kg, and packed in the carton (930 x 1,380 x 930 mm) around 189 kg. As with the smaller model, a door interlock for alternating opening can be ordered from the factory, as contamination protection for wall installation.

Control: ControlCOCKPIT TwinDISPLAY

The controller works with two high-resolution TFT colour graphic displays, and the HeatBALANCE function adjusts the heating power distribution between the upper and lower heating group in the range from -50 to +50 %. SetpointWAIT lets the process time start only when the setpoint temperature has actually been reached, and a calibration function covers three freely selectable temperature values. The electronically controllable air turbine works in 10 % steps, and an exhaust air flap admixes preheated fresh air. Programmes are managed via Ethernet or USB with AtmoCONTROL, and the sequence remains stored in the event of a power failure. The setting accuracy is 0.1 °C up to a setpoint temperature of 99.9 °C and 0.5 °C from 100 °C.

Electrical connection and installation

The UF260TS is connected at 230 V at approx. 3,400 W or at 115 V at approx. 1,800 W, in each case 50/60 Hz. An installation altitude of up to 2,000 m above sea level is permissible at an ambient temperature of +5 °C to +40 °C and a relative humidity of at most 80 %, non-condensing. The cabinet may only be transported upright.

Safety according to DIN 12880

Two Pt100 sensors of DIN class A monitor each other in 4-wire measurement and automatically take over the function of the other in the event of a fault. A mechanical temperature limiter of protection class 1 according to DIN 12880 switches off the heating around 20 °C above the maximum cabinet temperature, and AutoSAFETY additionally monitors over- and under-temperature with an alarm. A self-diagnosis system detects faults early, and visual and audible alarms report faults immediately.

Selection guide: between UF160TS and UF450TS

If the 161 litres of the smaller Memmert pass-through oven UF160TS are no longer sufficient, the UF260TS with 256 litres and up to 9 shelves is the obvious next size. For heavier goods of over 20 kg per shelf or a three-phase connection, the larger Memmert pass-through oven UF450TS is the more suitable choice. Without an airlock function and with access from one side, the likewise 256 litre Memmert universal oven UF260plus with TwinDISPLAY is often sufficient too.

Typical applications

The UF260TS is also designed for the direct transfer of load material and samples between the grey area and the clean room and reduces the risk of contamination when transporting the samples.

  • Airlock transfer of larger batches of load material and samples between the grey area and the clean room
  • Time saved when loading through direct access from both sides
  • Drying and annealing of samples with subsequent transfer into the clean room
  • Reducing the risk of contamination during sample transport through the double door function

Technical details

Setting temperature range+20 to +250 °C
Working temperature rangemin. 10 °C above room temperature to +250 °C
Setting accuracyup to 99.9 °C: 0.1 °C, from 100 °C: 0.5 °C
Display resolution, actual value0.1 °C
Temperature sensor2x Pt100, DIN class A, 4-wire measurement, redundant
TimerDigital countdown timer with target time, 1 min to 99 days
Internal volume256 l
Internal dimensions (W x H x D)640 x 800 x 500 mm
External dimensions (W x H x D)825 x 1,314 x 682 mm (plus 2x56 mm door handle)
Net weightapprox. 138 kg
Gross weight (carton)approx. 189 kg
Dimensions incl. carton930 x 1,380 x 930 mm
Connection 230 V50/60 Hz, approx. 3,400 W
Connection 115 V50/60 Hz, approx. 1,800 W
Shelves as standard / maximum2 / 9
Max. load, device300 kg
Max. load per shelf20 kg
ControlControlCOCKPIT TwinDISPLAY, PID microprocessor controller, 2x TFT colour graphic display
Additional functionsHeatBALANCE, SetpointWAIT, AutoSAFETY
Permissible ambient temperature+5 °C to +40 °C
Permissible humiditymax. 80 % rel., non-condensing
Maximum installation altitude2,000 m above sea level
Electrical categoryOvervoltage category II, pollution degree 2
MarkingsCE, EAC

Other sizes in the series

The UF260TS is one of five sizes of the TS series, all with TwinDISPLAY, airlock function and a setting range of +20 to +250 °C.

ModelInternal volumeSetting temperatureDisplay
UF160TS161 l+20 to +250 °CTwinDISPLAY
UF260TS (this page)256 l+20 to +250 °CTwinDISPLAY
UF450TS449 l+20 to +250 °CTwinDISPLAY
UF750TS749 l+20 to +250 °CTwinDISPLAY
UF1250TS (on request)1,248 l+20 to +250 °CTwinDISPLAY

Scope of delivery

  • Pass-through oven UF260TS with fully insulated stainless steel doors on both sides
  • Factory calibration certificate (calibration at +160 °C)
  • 2x stainless steel grid shelf, electropolished
  • Installation on feet

Regulatory information

The device is an electrical device within the meaning of the German Electrical and Electronic Equipment Act (ElektroG). The WEEE registration number of the manufacturer Memmert GmbH + Co. KG is DE 66812464. Information on returning old appliances can be found on our page Environment and disposal.

Installation in a wall opening

For fixed installation in a wall opening, Memmert offers a flush-fit unit set without air slots (B33205). It covers the gap between the pass-through oven and the wall opening with stainless steel. According to Memmert, the installation situation must be clarified technically beforehand. The set is therefore available on request through our procurement service.

Individual requirements and procurement service

For the UF260TS, options available include a full-view glass door with 4-layer insulating glazing up to 250 °C, a 115 V voltage variant, a door interlock for alternating opening and an additional Pt100 sensor. In addition, reinforced stainless steel grid shelves, slide-in and base drip trays, a built-in frame set for wall installation or a USB-Ethernet adapter can be ordered. For qualification and maintenance, we arrange factory and DAkkS calibration, IQ/OQ documentation according to DIN 12880:2007-05 as well as instruction and maintenance contracts. Our procurement service supports you with individual requirements and with selecting the right equipment.

Volume: 256 l
Skip product gallery

Matching accessories

Memmert Edelstahl-Gitterrost, Zeichnung verschiedener Größen
Memmert stainless steel grid E28891 for sizes 260 and 410
Manufacturer number: E28891

The E28891 stainless steel grid is an electropolished shelf for 26 Memmert models in sizes 260 and 410. Each grid is a shelf level for the items loaded into the interior. If you need more shelf levels than are fitted as standard, you can add grids up to the maximum number in the table. The size corresponds to the number in the unit name, for example UN260 or IF260.All 26 models have 2 grids as standard. The maximum number is 9 grids (size 260) and 14 grids (size 410). According to Memmert, the load per shelf is 20 kg. The image is for illustration only. It is a Memmert drawing showing grids of different sizes.Typical applicationsTypical uses in the compatible units:Drying several sample series in parallel in routine operation, for example in the UF260Incubating extensive sample series in research and medicine, for example in the IF260Parallel stability studies of several active substance batches according to ICH Q1A, for example in the HPP260ecoMicrobiological examinations and microbial count determination in the routine laboratory, for example in the IPP410ecoSuitable forThe table shows for each model the number of grids included in the scope of delivery of the unit, the maximum possible number and the load per shelf.ModelGrids as standardMaximum number of gridsLoad per shelf (kg)Interior volume (l)UN260plus2920256UN2602920256IF260plus2920256IN260plus2920256IF2602920256IN2602920256UF2602920256UF260plus2920256IF260bw2920256IPP260eco2920256IPP260ecoPlus2920256IPP410eco21420384IPP410ecoplus21420384HPP260eco2920256HPP410eco21420384ICH2602920256ICP2602920256IN260m2920256IN260mplus2920256IF260m2920256IF260mplus2920256UN260m2920256UN260mplus2920256UF260m2920256UF260mplus2920256UF260TS2920256The permissible total load of the unit additionally limits the loading. Please refer to the information in the unit’s data sheet.Technical detailsMaterialStainless steelSurfaceelectropolishedLoad per shelf according to Memmert20 kgScope of delivery1 stainless steel grid, electropolished (E28891)Individual requirements & procurement serviceFor further grids, perforated trays or reinforced grids, and for questions about compatibility with units not listed in the table, our procurement service accepts enquiries.

Manufacturer's RRP: €126.42
memmert einschiebe abtropfschale
Memmert stainless steel slide-in drip tray B29726 for sizes 260 and 410
Manufacturer number: B29726

The stainless steel slide-in drip tray B29726 is intended for Memmert units in the U, Um, UF TS, I, Im, IPPeco, ICP, ICH and HPPeco series in sizes 260 and 410. It has a 15 mm rim and collects dripping liquid. Memmert states a maximum load of 4 kg per slide-in drip tray. The size corresponds to the number in the unit name, for example UN260 or IF260.According to Memmert, it can impair the spatial temperature distribution. The table of compatible models shows which series is available in which size. The base drip tray for the same unit size is also available (B29722). The image is for illustration only.Typical applicationsTypical uses in the compatible units:Forcing tests for beer stability, for example in the IPP410ecoShelf-life tests on dairy products, for example in the IPP410ecoSuitable forModelSizeInterior volume (l)UN260plus260256UN260260256IF260plus260256IN260plus260256IF260260256IN260260256UF260260256UF260plus260256IPP260eco260256IPP260ecoPlus260256IPP410eco410384IPP410ecoplus410384HPP260eco260256HPP410eco410384ICH260260256ICP260260256IN260m260256IN260mplus260256IF260m260256IF260mplus260256UN260m260256UN260mplus260256UF260m260256UF260mplus260256UF260TS260256Technical detailsMaterialStainless steelRim15 mmMaximum load according to Memmert4 kgScope of delivery1 stainless steel slide-in drip tray, 15 mm rimIndividual requirements & procurement serviceFor the base drip tray, for grids and perforated trays of the same size, and for questions about compatibility with older Memmert units, our procurement service accepts enquiries.

Manufacturer's RRP: €164.26
memmert boden abtropfschale
Memmert stainless steel base drip tray B29722 for sizes 260 and 410
Manufacturer number: B29722

B29722 is a stainless steel base drip tray with a 15 mm rim for Memmert units in the U, Um, UF TS, I, Im, IPPeco, ICP, ICH and HPPeco series in sizes 260 and 410. It collects dripping liquid. According to Memmert, the maximum load is 4 kg per base drip tray. The size corresponds to the number in the unit name, for example UN260 or IF260.Memmert points out that it can impair the spatial temperature distribution. The table of compatible models shows which series is available in which size. The slide-in drip tray for the same unit size is also available (B29726). The image is for illustration only.Typical applicationsTypical uses in the compatible units:Forcing tests for beer stability, for example in the IPP410ecoShelf-life tests on dairy products, for example in the IPP410ecoSuitable forModelSizeInterior volume (l)UN260plus260256UN260260256IF260plus260256IN260plus260256IF260260256IN260260256UF260260256UF260plus260256IPP260eco260256IPP260ecoPlus260256IPP410eco410384IPP410ecoplus410384HPP260eco260256HPP410eco410384ICH260260256ICP260260256IN260m260256IN260mplus260256IF260m260256IF260mplus260256UN260m260256UN260mplus260256UF260m260256UF260mplus260256UF260TS260256Technical detailsMaterialStainless steelRim15 mmMaximum load according to Memmert4 kgScope of delivery1 stainless steel base drip tray, 15 mm rimIndividual requirements & procurement serviceFor the slide-in drip tray, for grids and perforated trays of the same size, and for questions about compatibility with older Memmert units, our procurement service accepts enquiries.

Manufacturer's RRP: €209.84
Stainless steel perforated shelf, illustration
Memmert stainless steel perforated tray B68510 for sizes 260 and 410
Manufacturer number: B68510

The B68510 stainless steel perforated tray is intended for Memmert units in the U, Um, UF TS, I, Im, IPPeco, ICP, ICH, HPPeco and IFbw series in sizes 260 and 410. It is a shelf and therefore one shelf level in the interior. The size corresponds to the number in the unit name, for example UN260 or IF260. The image is for illustration only.According to Memmert, the maximum load per shelf for these sizes is 20 kg. In addition, the maximum load per unit limits the loading; it is stated in the unit’s data sheet. The table of compatible models shows which series is available in which size.Typical applicationsTypical uses in the compatible units:Drying several sample series in parallel in routine operation, for example in the UF260Incubating extensive sample series in research and medicine, for example in the IF260Parallel stability studies of several active substance batches according to ICH Q1A, for example in the HPP260ecoMicrobiological examinations and microbial count determination in the routine laboratory, for example in the IPP410ecoSuitable forModelSizeInterior volume (l)UN260plus260256UN260260256IF260plus260256IN260plus260256IF260260256IN260260256UF260260256UF260plus260256IF260bw260256IPP260eco260256IPP260ecoPlus260256IPP410eco410384IPP410ecoplus410384HPP260eco260256HPP410eco410384ICH260260256ICP260260256IN260m260256IN260mplus260256IF260m260256IF260mplus260256UN260m260256UN260mplus260256UF260m260256UF260mplus260256UF260TS260256Technical detailsMaterialStainless steelLoad per shelf according to Memmert20 kgScope of delivery1 stainless steel perforated trayIndividual requirements & procurement serviceFor grids, further perforated trays and drip trays, and for questions about compatibility with older Memmert units, our procurement service accepts enquiries.

Manufacturer's RRP: €206.40