Skip to main content Skip to search Skip to main navigation

Memmert Incubator IF55mplus, 53 l, +20 to +80 °C, TwinDISPLAY

Product information "Memmert Incubator IF55mplus, 53 l, +20 to +80 °C, TwinDISPLAY"

The Memmert Incubator Im IF55mplus is a class I medical device under Regulation (EU) 2017/745, with the intended purpose of warming irrigation and infusion solutions and contrast media. The 53 litre chamber lies between the smallest and the middle model of the mplus series and is suitable for smaller quantities. A stainless steel grid shelf is included as standard, and up to four shelves are possible.

Control, HeatBALANCE and program transfer

Two mutually monitoring Pt100 sensors of DIN class A work in 4-wire measurement and, in the event of a fault, automatically take over the function of the other. The setting accuracy is 0.1 °C with a setting temperature range of +20 to +80 °C. The ControlCOCKPIT TwinDISPLAY with two TFT colour graphic displays controls the cabinet, and the HeatBALANCE function adjusts the heating power distribution between the upper and lower heating group in the range from -50 to +50 per cent. Via the Ethernet interface or USB port, programs can be created, managed and transferred with the AtmoCONTROL software, and the program sequence is stored in the event of a power failure. The SetpointWAIT function only starts the process time once the setpoint temperature has been reached, and three temperature values can be freely calibrated.

Safety technology

The mechanical temperature limiter of protection class 1 to DIN 12880 switches the heating off at approx. 20 °C above the maximum cabinet temperature. In addition, an adjustable temperature monitor of protection class 3.1 or an adjustable temperature limiter of protection class 2 can be selected on the display. The AutoSAFETY monitoring follows the setpoint at a freely selectable distance, reports deviations visually and audibly, and switches the heating off when exceeded. A self-diagnosis system helps with fault finding. A fully insulated stainless steel door with 2-point locking (compression door seal) and an inner glass door close the chamber, and the stainless steel interior is stiffened by deep-drawn ribbing and has protected, integrated 4-sided all-round heating. The permanently installed sterilisation program (4 h at 160 °C) is used for sterilising the interior and is not intended for sterilising the load.

Ventilation, connection and dimensions

The air turbine is electronically adjustable in 10 % steps and runs quietly, and preheated fresh air is admixed via the electronically adjustable exhaust air flap. The IF55mplus is operated with 230 V at approx. 1,000 W or with 115 V at approx. 900 W, each 50/60 Hz. The internal dimensions are 400 x 400 x 330 mm, the housing measures 585 x 784 x 514 mm externally, and the net weight is around 57 kg. A total of up to 80 kg and 20 kg per shelf may be loaded.

Distinction from the SingleDISPLAY variant and the laboratory device

We carry the IF55m as the SingleDISPLAY variant with only one Pt100 sensor, also a class I medical device, but with a different intended purpose: warming and keeping warm of fango, silicate and APS packs in physiotherapy. The IF55mplus differs from it in the TwinDISPLAY, the two redundant sensors, HeatBALANCE, AutoSAFETY and the adjustable temperature monitor or limiter selectable on the display. Without a medical intended purpose, but with technically comparable equipment, the IF55plus is also available as a laboratory incubator of the same chamber size.

Selection guide: smaller or larger chamber

Those who can manage with less volume will find the smallest size in the series in the IF30mplus with 32 litres. If 53 litres are not enough, the IF75mplus with 74 litres and up to six shelves offers more capacity.

Typical applications

The IF55mplus is intended, as a class I medical device, for the following applications:

  • Warming irrigation solutions
  • Warming infusion solutions
  • Warming contrast media

Technical details

Setting temperature range+20 to +80 °C
Working temperature rangemin. 10 °C above room temperature to +80 °C
Temperature setting accuracy0.1 °C
Temperature sensor2x Pt100, DIN class A, 4-wire measurement, redundant
Internal volume53 l
Internal dimensions (W x H x D)400 x 400 x 330 mm
External dimensions of housing (W x H x D)585 x 784 x 514 mm
Net weightapprox. 57 kg
Connection230 V, 50/60 Hz, approx. 1,000 W; 115 V, 50/60 Hz, approx. 900 W
Shelves in scope of delivery1
Maximum shelves4
Max. load of unit80 kg
Max. load per shelf20 kg
ControlControlCOCKPIT TwinDISPLAY, PID microprocessor controller, 2x TFT colour graphic display
Additional functionsHeatBALANCE, AutoSAFETY, SetpointWAIT, programming via AtmoCONTROL (Ethernet/USB)
Medical device classificationClass I under Regulation (EU) 2017/745
Intended purposeWarming of irrigation and infusion solutions and contrast media
Permissible ambient temperature+5 °C to +40 °C
Permissible humiditymax. 80 % rel., non-condensing
MarkingsCE, VDE, cULus

Other sizes in the series

The mplus series offers eight chamber sizes from 32 to 749 litres, each with TwinDISPLAY and the setting temperature range +20 to +80 °C. With 53 litres, the IF55mplus is in second place.

ModelInternal volumeSetting temperatureDisplay
IF30mplus32 l+20 to +80 °CTwinDISPLAY
IF55mplus (this page)53 l+20 to +80 °CTwinDISPLAY
IF75mplus74 l+20 to +80 °CTwinDISPLAY
IF110mplus108 l+20 to +80 °CTwinDISPLAY
IF160mplus161 l+20 to +80 °CTwinDISPLAY
IF260mplus256 l+20 to +80 °CTwinDISPLAY
IF450mplus449 l+20 to +80 °CTwinDISPLAY
IF750mplus749 l+20 to +80 °CTwinDISPLAY

Scope of delivery

  • 1x Memmert Incubator Im IF55mplus
  • Factory calibration certificate at +37 °C
  • Fully insulated stainless steel door with 2-point locking (compression door seal)
  • Inner glass door
  • 1x stainless steel grid shelf, electropolished

Regulatory information

The device is electrical equipment within the meaning of the German Electrical and Electronic Equipment Act (ElektroG). The WEEE registration number of the manufacturer Memmert GmbH + Co. KG is DE 66812464. Notes on returning old devices can be found on our page Environment and disposal. The IF55mplus also carries the classification as a class I medical device under Regulation (EU) 2017/745; the intended purpose includes warming irrigation and infusion solutions and contrast media.

Individual requirements and procurement service

For the IF55mplus, the 115 V voltage variant and temperature limiting are available as factory options. For accessories, we arrange grid shelves, perforated trays, drip trays for shelf or floor, a wall bracket, a stacking kit or a levelling or mobile base. For DAkkS calibration, IQ/OQ/PQ documentation and qualification or a maintenance contract, we arrange the right contact. Contact us for individual advice.

Volume: 53 l