Memmert Incubator IF750mplus, 749 l, +20 to +80 °C, TwinDISPLAY
€8,254.00
Available, delivery time: 2-3 weeks
- Class I medical device under Regulation (EU) 2017/745
- Intended purpose: tempering of irrigation and infusion solutions and contrast media
- 749 l internal volume, TwinDISPLAY with 2 TFT colour graphic displays
- HeatBALANCE: adjustable heating power distribution -50 % to +50 %
- 2x redundant Pt100 sensors with function takeover in the event of a fault
- AutoSAFETY monitoring with visual and audible alarm
- Factory calibration certificate included (calibration at +37 °C)
Product information "Memmert Incubator IF750mplus, 749 l, +20 to +80 °C, TwinDISPLAY"
With 749 litres and up to 14 shelves, the Memmert Incubator Im IF750mplus forms the top end of the mplus series. Its intended purpose as a Class I medical device under Regulation (EU) 2017/745 is the tempering of irrigation and infusion solutions and contrast media. Two stainless steel grid shelves are included as standard.
Pt100 pair with function takeover
Two Pt100 sensors of DIN class A measure using 4-wire technology and monitor each other; if one fails, the other takes over the measuring function. A temperature between +20 and +80 °C can be set, and the setting accuracy is 0.1 °C. The display is provided by the ControlCOCKPIT TwinDISPLAY with two TFT colour graphic displays.
Heating power and program transfer
HeatBALANCE shifts the heating power between the upper and lower heating group in the range of -50 to +50 percent. AtmoCONTROL via Ethernet or USB is used for programming, the sequence remains stored in the event of a power failure, and SetpointWAIT lets the process time begin only once the set temperature has been reached. Three values can be freely calibrated.
Limiters, monitors and sterilisation
A mechanical temperature limiter of protection class 1 to DIN 12880 switches the heating off at approx. 20 °C above the maximum cabinet temperature. Optionally, an adjustable temperature monitor of protection class 3.1 or an adjustable temperature limiter of protection class 2 can be set on the display. A permanently installed sterilisation program (4 h at 160 °C) is used for sterilising the interior and is not intended for sterilising load items. A self-diagnosis system additionally assists with fault finding.
AutoSAFETY monitoring
In addition to the temperature limiter, AutoSAFETY automatically follows the set value within a freely selectable range above and below. If the temperature deviates, the device triggers a visual and audible alarm, and if the value is exceeded, it disconnects the heating.
Dimensions, connection and installation
The housing stands mobile and lockable on castors. The IF750mplus is connected at 230 V with approx. 2,000 W or at 115 V with approx. 1,800 W. Internally, the cabinet measures 1,040 x 1,200 x 600 mm and externally 1,224 x 1,720 x 784 mm, with a net weight of approx. 217 kg, and a total load of 300 kg is permitted at 30 kg per shelf, distributed over up to 14 shelves.
Comparison with IF750m and IF750plus
With the IF750m, a single Pt100 sensor handles the measurement, controlled via SingleDISPLAY without HeatBALANCE, and the intended purpose remains a different one (fango, silicate and APS packs). The laboratory incubator IF750plus also has a TwinDISPLAY, but no medical intended purpose.
Largest chamber in the series
The IF750mplus is the largest chamber in the mplus series, and Memmert does not offer a larger size in this line. If 449 litres are sufficient, the smaller, also mobile IF450mplus is the more compact alternative.
Typical applications
The intended purpose of the IF750mplus as a Class I medical device includes:
- Tempering of irrigation solutions
- Tempering of infusion solutions
- Tempering of contrast media
Technical details
| Setting temperature range | +20 to +80 °C |
| Working temperature range | min. 10 °C above room temperature to +80 °C |
| Temperature setting accuracy | 0.1 °C |
| Temperature sensor | 2x Pt100, DIN class A, 4-wire measurement, redundant |
| Internal volume | 749 l |
| Internal dimensions (W x H x D) | 1,040 x 1,200 x 600 mm |
| External dimensions of housing (W x H x D) | 1,224 x 1,720 x 784 mm |
| Net weight | approx. 217 kg |
| Connection | 230 V, 50/60 Hz, approx. 2,000 W; 115 V, 50/60 Hz, approx. 1,800 W |
| Shelves in scope of delivery | 2 |
| Shelves, maximum | 14 |
| Max. load, unit | 300 kg |
| Max. load per shelf | 30 kg |
| Installation | mobile and lockable, on castors |
| Control | ControlCOCKPIT TwinDISPLAY, PID microprocessor controller, 2x TFT colour graphic display |
| Additional functions | HeatBALANCE, AutoSAFETY, SetpointWAIT, programming via AtmoCONTROL (Ethernet/USB) |
| Medical device classification | Class I under Regulation (EU) 2017/745 |
| Intended purpose | Tempering of irrigation and infusion solutions and contrast media |
| Permissible ambient temperature | +5 °C to +40 °C |
| Permissible humidity | max. 80 % rel., non-condensing |
| Markings | CE, VDE, cULus |
Other sizes in the series
The mplus series comprises eight chamber sizes from 32 to 749 litres, all with TwinDISPLAY and the setting temperature range +20 to +80 °C. At 749 litres, the IF750mplus forms the top end.
| Model | Internal volume | Setting temperature | Display |
|---|---|---|---|
| IF30mplus | 32 l | +20 to +80 °C | TwinDISPLAY |
| IF55mplus | 53 l | +20 to +80 °C | TwinDISPLAY |
| IF75mplus | 74 l | +20 to +80 °C | TwinDISPLAY |
| IF110mplus | 108 l | +20 to +80 °C | TwinDISPLAY |
| IF160mplus | 161 l | +20 to +80 °C | TwinDISPLAY |
| IF260mplus | 256 l | +20 to +80 °C | TwinDISPLAY |
| IF450mplus | 449 l | +20 to +80 °C | TwinDISPLAY |
| IF750mplus (this page) | 749 l | +20 to +80 °C | TwinDISPLAY |
Scope of delivery
- 1x Memmert Incubator Im IF750mplus
- Factory calibration certificate at +37 °C
- Fully insulated stainless steel doors with 2-point locking (compression door seal)
- Inner glass doors
- 2x stainless steel grid shelf, electropolished
Regulatory information
The device is electrical equipment within the meaning of the German Electrical and Electronic Equipment Act (ElektroG). The WEEE registration number of the manufacturer Memmert GmbH + Co. KG is DE 66812464. Information on the return of old equipment can be found on our page Environment and disposal. The IF750mplus is also classified as a Class I medical device under Regulation (EU) 2017/745; its intended purpose is the tempering of irrigation and infusion solutions and contrast media.
Individual requirements and procurement service
The 115 V voltage variant and temperature throttling are available as factory options for the IF750mplus. For accessories, we source grid shelves, perforated trays, insert and base drip trays, a built-in frame, and a USB-Ethernet adapter or an Ethernet connection cable. For DAkkS calibration, IQ/OQ/PQ documentation and qualification as well as maintenance contracts, we put you in touch with the right contact. Contact us for individual advice.
| Volume: | 749 l |
|---|